16
3. Press again, the display will show ‘01’, which means the latest reading, then turns to another screen
to show readings and measuring time as Fig.15;
4. Press again, the display will show ‘02’, which means the second to the latest reading…
Memory clearance
After a measurement is finished or when the device stands by, hold down button ‘M1’ or
‘M2’ for at least 5 seconds, the display will show ‘CLR’ which means the stored reading
for ‘M1’ or ‘M2’ is removed.
IRREGULAR HEARTBEAT DETECTOR
Model Domino digital blood pressure monitor provides a blood pressure and pulse rate
measurement even when an irregular heartbeat occurs. When the device detects the
irregular heartbeat or any excessive body movement during measurement, the ‘IHB ’
icon will display in the LCD. It is important that you be relaxed, remain still and do not
talk during measurement.
Note: We recommend contacting your physician if you see this ‘IHB’ indicator frequent-
ly.
ERROR AND LOW BATTERY INFORMATION
CARE, STORING, REPAIR AND RECYCLING
1. It’s necessary to protect this device against high moisture, direct sunlight, shock, solvent, alcohol and
gasoline.
2. Remove the batteries if the device is to be stored for a long time, and keep the batteries far from the
children.
3. Keep the cuff from sharp subject and don’t extend or twist the cuff.
4. Use only soft and dry cloth to clean the device.
5. The cuffs are sensitive and must be handled with care. You can clean the cuff cover with damp cloth.
WARNING: Under no circumstances may you wash the inner bladder!
6. It is necessary to consult specialists yearly for checking technical condition of the device. Please
consult your dealer for more information.
INDICATION
POSSIBLE REASON
CORRECTION METHODS
The cuff is put on wrongly or the tube
plug is inserted too loosely.
Make sure that cuff is put on correctly
and the tube plug is inserted tightly and
repeat the measurement.
Movement of arm/hand or talking during
measurement.
Repeat the measurement with following
completely recommendations of ma-
nual.
The cuff is not inflated to necessary
pressure.
Repeat the measurement with pumping
cuff on 30-40 mmHg above expected
systolic pressure.
The batteries are weak.
Replace all 4 batteries with new ones.