19
MAINTENANCE
The Mistral device does not have any part that requires maintenance and/or lubrication.
Nonetheless, it is necessary to carry out some checks to verify the function and security of
the appliance before each use. Remove the appliance from the box and
always check
that
there is no visible damage; particular attention should be given to any cracks in plastics
which may leave some electrical components exposed. Check also the integrity of the pow-
er cable which might have been damaged during the previous use.
Then connect the cable to the power network and switch it on. Close the compression cap
with one finger and check that there are no disturbing noises which may be evidence of
malfunction.
Check that the nebulizer has not suffered any breakage during the previous use (it was
stored inappropriately or has suffered impacts). The appliance is protected by a protection
fuse
(F 2A L 250V)
not accessible from outside. Therefore, contact the technical staff au-
thorized by the manufacturer for its replacement.
If the appliance still does not nebulize after having checked the above mentioned condi-
tions, it is recommended to contact the reseller or Gima’s technical service.
BEFORE CARRYING OUT ANY CHECKS IN THE EVENT OF ANOMALIES OR
MALFUNCTIONS, CONTACT GIMA’S TECHNICAL SERVICE. THE MANUFACTURER
OFFERS NO GUARANTEE FOR THE APPLIANCES WHICH HAVE ASSESSED AS TAM-
PERED AFTER THE CHECKS CARRIED OUT BY THE TECHNICAL SERVICE.
CLEANING AND WASHING OF ACCESSORIES
Before each use and / or after the cleaning operations, check the integrity of all accessories
supplied with the device. Turn off the device before each cleaning operation and disconnect
the power cable from the socket.
Type of defect
Cause
Corrective action
1. Poor atomization. Clogged ampoule.
Clean and disinfect the ampoule as set
out in the manual.
2. Poor atomization. Clogged ampoule.
If the washing had no positive out-
come, replace the ampoule.
3. Lack of
atomization.
Nozzle stuck.
Press hard the nozzle (cylindrical im-
mersion tube) placed inside the poly-
carbonate ampoule bottom with a finger.
4. Slow atomization. Excessive oily drug.
Dilute the drug with saline.
5. Noisy device.
Prolonged use.
Contact the reseller or Gima’s technical
service.
6. The Device
does not work.
Defective power cable
Broken and/or absent
power source.
Replace the power cable.
Check the power source and voltage.
Defects 1 - 2 - 3 -
4 - 5 - 6
None of the corrective
actions has been effective
Contact the reseller or Gima’s technical
service.