The tables below provide information on the EMC characteristics (Electromagnetic Compat-
ibility) of this electromedical appliance.
ENGLISH
Guide and declaration of the manufacturer – Electromagnetic Emissions
The Corsia aerosol can be used in the following electromagnetic environment. The Cus-
tomer and/or the user of the Corsia aerosol shall make sure that the appliance is used in
such environment.
Emission test
Conformity Guide to the electromagnetic
environment
Radiated emissions /
Conductions CISPR11
Group 1
The Corsia aerosol uses RF energy only for
Internal function. Therefore, its RF emissions
are very low and do not cause any interfer-
ence with other nearby electronic applianc-
es.
Radiated emissions /
Conductions CISPR11
Class [B]
The Corsia aerosol is designed for use in
any environment, including households and
those directly connected to the public power
distribution grid which supplies power to en-
vironments intended for domestic use.
Harmonic currents EN
61000-3-2
Class [A]
Voltage fluctuations /
flicker EN 61000-3-3
Compliant
Guide and declaration of the manufacturer – Electromagnetic Immunity
The Corsia aerosol can be used in the following electromagnetic environment. The Cus-
tomer and/or the user of the Corsia aerosol shall make sure that the appliance is used in
such environment.
Immunity test
Standard of proof
Level of
conformity
Guide to the electromag-
netic environment
Electrostatic
discharge (ESD)
EN 61000-4-2
± 8kV contact
± 15kV air
The appliance
does not alter
its status
Floors should be in wood,
cement or ceramics. If floors
are covered by synthetic ma-
terial, the relative humidity
should be at least 30%.
Fast transient / burst
EN 61000-4-4
± 2kV power supply
± 1kV signal cables
The appliance
does not alter
its status
The power supply should be
that of a typical commercial
premise or hospital.
Surge EN 61000-4-5 ± 0.5kV e± 1kV
differential mode
The appliance
does not alter
its status
The power supply should be
that of a typical commercial
premise or hospital.
Voltage dips, short
outages and
voltage variations
EN 61000-4-11
5%
U
T
(>95% dip
in
U
T
) per 0.5 cycle
40%
U
T
(60% dip
in
U
T
) per 5 cycles
70%
U
T
(30% dip
in
U
T
) per 25 cycles
<5%
U
T
(>95% dip
in
U
T
) per 5 s
--
The power supply should be
that of a typical commercial
premise or hospital.
If the user of the Corsia aer-
osol requires the appliance to
work continuously, it is rec-
ommended to use it with an
uninterruptible power supply.