Attention: not complying with warnings will relieve GIMA company 

 

from any kind of responsibility.

Chapter 2

INSTALLATION

2.1 Trasport and stocking

Before shipping the podoscope 27363 is covered by a film and packed 

in a cardboard with protective panels inside.

The box contains the assembled podoscope and its cable (which must be 

obviously connected to the apparatus before using it); the caoutchouc 

is not included. 

In case of a longer storage, please store the whole package clean and dry.

2.2 Conditions for installation

1. Unpack the podoscope completely.

2. Disposal of packing elements (plastic foams, polystyrene, polyethylene,

  cardboard) must follow your local waste disposal guidelines.

3. Verify that the podoscope has not been damaged during transport. In case, 

  please contact the manufacturer immediately.

IMPORTANT

 Please take notice of the fact that light stripes and/or 

microscopic luminescent dots on the transparent surface of the podoscope 

are due to the particular nature of its material and do not impair 

the equipment functionality.

RIGHT POSITIONING AND SAFETY INSTALLATION OF THE PODOSCOPE

Position the podoscope so as to leave enough room around it to be able to 

safely operate and to connect and disconnect easily the device to the power 

supply network.

We suggest to attach a handle on the wall as indicated in the scheme below 

to make easier for patients to go on and down the podoscope.

ENGLISH

Содержание 27363

Страница 1: ...efore using the product AVIS Les op rateurs doivent lire et bien comprendre ce manuel avant d utiliser le produit ATENCI N Los operadores tienen que leer y entender completamente este manual antes de...

Страница 2: ...ence norms 17 1 5 Electrical features and voltage 18 1 6 Components identification 18 Electrical schema 19 1 7 General safety warning 19 Chapter 2 INSTALLATION 20 2 1 Transport and stocking 20 2 2 Con...

Страница 3: ...USAGE MAINTENANCE Instructions for a correct installation are contained in the par 2 2 CONDITIONS FOR THE INSTALLATION Follow the instructions for use Read carefully this manual for the installation...

Страница 4: ...an discover variations of colour intensity on the foot plant corresponding to higher or lower plant pressure The base is of ABS A separable net cable with plug a jack and a led strip are its electric...

Страница 5: ...sports centres by operators with knowledge of biomechanics like doctors in sport medicine orthopedic specialists and physiotherapists Reference harmonized norms CEI EN 60601 1 Safety of electromedical...

Страница 6: ...assembled and fixed in such a solid way so as to stand up to the often adverse transport conditions As to indirect shocks the device is classified in Class II As to current leakage it can be classifi...

Страница 7: ...always take the following precautions A Always unplug the podoscope during maintenance operations B Do not move the podoscope while using it C Each time before the usage check the functioning of the c...

Страница 8: ...ystyrene polyethylene cardboard must follow your local waste disposal guidelines 3 Verify that the podoscope has not been damaged during transport In case please contact the manufacturer immediately I...

Страница 9: ...00 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Compliance Group1 Class B Compliant Compliant Electromagnetic environment guidance The equipment 27363 must emit electromagnetic energy in o...

Страница 10: ...ment Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of 27363 requires...

Страница 11: ...the examination can be performed only if the patient is barefoot otherwise lay the caoutchouc on the transparent surface This material is also useful to check the validity of arch support treatments W...

Страница 12: ...19 Then get out the blown fuse with a suitable instrument and replace it with a T 0 5 A 5x20 250V fuse If you have to replace the power cord the new one must have the same characteristics and certifi...

Страница 13: ...he warranty GIMA will repair and or replace free of charge all the defected parts due to production reasons Labor costs and personnel traveling expenses and packaging not included All components subje...

Страница 14: ...74...

Страница 15: ...75...

Страница 16: ......

Отзывы: