leave the thermometer unattended in the hands of children.
• Avoid bending the thermometer tip.
• If the ambient temperature is over 35°C or 95°F, dip the thermome-
ter tip in cold water for approx. 5 to 10 seconds prior to measuring
the temperature. Persistent fever, in particular in children, has to be
treated by a doctor - please get in touch with your doctor!
• Do not use near strong electromagnetic fields, i.e.
Keep it away from any radio systems and mobile phones.
Battery replacement
The battery is empty and needs replacing when the
or
battery symbol appears on the right of the LCD. Remove the battery
cover and replace it with a battery (preferably non-mercury) of the
same type.
We advise you to remove the batteries if the device is not going to be
used for a longer period of time.
Technical data
Type: maximum thermometer
Measurement range: (32.0~ 42.9)°C/(89.6~109.2)°F
Measurement accuracy:
Measurement accuracy:
+/- 0.1°C/0.2°F (35.5°C~42.0°C/95.9°F~107.6°F)
+/-0.2°C/0.4°F (32.0°C~35.5°C/89.6°F~95.9°F)
(42.0°C~42.9°C/107.6°F~109.2°F)
Storage/transportation temperature: (-25~55)°C, ≤95%RH
Ambient temperature during use: (5~35)°C, ≤80%RH
Atmosferic pressure 700~1060 hPa
Min Scale: 0.1°C/0.1°F
Battery type: alkaline battery, type LR41, 1.5V, service life minimum
100 hours under continuous operation.
Explanation of symbols
Legal requirements and guidelines
This product complies with the European Directive for Medical Devic-
es 93/42/EEC and carries the CE mark. The device also complies with
the specifications of the European Standard EN 12470-3 Clinical ther-
mometers-Part 3: Performance of compact electrical thermometers
(non-predictive and predictive) with maximum device.
Calibration check
This thermometer is initially calibrated at the time of manufacture. If
this thermometer is used according to the operation instruction, peri-
odic re-adjustment is not required.
The calibration check has to be carried out immediately, if there are
indications that the product does not keep the defined error limits or
the calibration properties could have been affected by an intervention
or by any other means. Please also observe any national statutory
regulations. The calibration check can be carried out by the compe-
tent authorities or by authorised service providers. A test instruction
for calibration check can be provided to the relevant authorities and
authorised services providers on request.
GIMA WARRANTY CONDITIONS
Congratulations for purchasing a GIMA product. This product meets
high qualitative standards both as regards the material and the pro-
duction. The warranty is valid for 12 months from the date of supply
of GIMA. During the period of validity of the warranty, GIMA will repair
and/or replace free of charge all the defected parts due to production
reasons. Labor costs and personnel traveling expenses and packag-
ing not included. All components subject to wear are not included in
the warranty. The repair or replacement performed during the warranty
period shall not extend the warranty. The warranty is void in the fol-
lowing cases: repairs performed by unauthorized personnel or with
non-original spare parts, defects caused by negligence or incorrect
use. GIMA cannot be held responsible for malfunctioning on electron-
ic devices or software due to outside agents such as: voltage chang-
es, electro-magnetic fields, radio interferences, etc. The warranty is
void if the above regulations are not observed and if the serial code (if
available) has been removed, cancelled or changed.
The defected products must be returned only to the dealer the prod-
uct was purchased from. Products sent to GIMA will be rejected.
Disposal:
The product must not be disposed of along with oth-
er domestic waste. The users must dispose of this equipment
by bringing it to a specific recycling point for electric and elec-
tronic equipment. For further information on recycling points
contact the local authorities, the local recycling center or the
shop where the product was purchased. If the equipment is
not disposed of correctly, fines or penalties may be applied in
accordance with the national legislation and regulations.
or
Battery is empty
WEEE disposal
Lo°C o Lo°F
Temperature under 32°C or 89.6°F
Hi°C o Hi°F
Temperature over 42.9°C or 109.2°F
Type BF applied part
Caution: read instructions (warnings) carefully
Keep away from sunlight
Keep in a cool, dry place
Follow instructions for use
Manufacturer
Medical Device complies with Directive 93/42/EEC
Product code
Lot number
IP22
Covering Protection rate