THE FAULT PHENOMENON
REASON
SOLUTION
Can't rise or down normally
1. The power cord is unconnected or not
plugged well
1. Connect the power cord or plug it well
2. Poor quality of the hand controller or foot
controller
2. Replace the hand controller or foot controller
3. Poor quality of the built-in foot controller
3. Please contact maintenance center or supplier for trouble-
shooting.
4. Poor quality of the rocker switch
5. Poor quality of the control box
6. Repture of motor tail
7. Motor was damaged
8. Pressing the button, the bed did not move 1. Put the movable footrest close to the fixed footrest, make
sure the limit switch is closed, and then press the buttons
again
2. Open the fixed footrest cover and backrest cover respec-
tively, check whether the two ends of the power cord are
loose, if the above situation occurs, please reconnect
3. If problem is still not solved after above operation, please
contact maintenance center or supplier to solve it.
Abnormal Sound
1. The rotating parts are short
of lubricating oil
1. Add lubricating oil
2. Rotating positions (such as hinge, POM
sets) are short of lubricating oil
2. Add lubricating oil
3. Motor quality problems
3. Replace the motor