PIPETMAN

®

 P

Dual-position adapter for P2 and P10

with D10

with DL10

with D10

Pre-rinse the tips

Some liquids (e.g. protein-containing solutions and 
organic solvents) can leave a film of liquid on the 
inside wall of the tip ; pre-rinse the tip to minimize any 
errors that may be related to this phenomenon. 
Pre-rinsing  consists  of  aspirating  the  first  volume 
of liquid and then dispensing it back into the same 
vessel (or to waste). Subsequent volumes that you 
pipette  will  have  levels  of  accuracy  and  precision 
within specifications.

aspirate

Press  the  push-button  to  the 

first  stop

  (this 

corresponds to the set volume of liquid).  
Hold the pipette vertically and immerse the tip 
in the liquid (see immersion depth table, page 8). 
Release the push-button slowly and smoothly (to 

top 

position) to aspirate the set volume of liquid. 

Wait one second (time depends on model, see 
table) ; then withdraw the pipette-tip from the 
liquid. You  may  wipe  any  droplets  away  from 
the  outside  of  the  tip  using  a  medical  wipe, 
however if you do so take care to avoid touching 
the tip’s orifice.

Содержание PIPETMAN P Series

Страница 1: ...EN ENGLISH User s Guide PIPETMAN P FR t l chargeable sur www gilson com auf www gilson com herunterladen se puede telecargar desde www gilson com scaricabile dal sito www gilson com DE SP IT...

Страница 2: ...r many applications P2 and P10 measurement and transfer of micro volumes DNA sequencing and enzyme assays P20 P100 P200 P1000 measurement and transfer of general aqueous solutions acids and bases Cont...

Страница 3: ...Safety bag Certi cate of conformity including bar code sticker 3 descrIPtIon Pleaserefertothefollowingchaptersforafulldescription of the different parts and functions of the pipette of the different p...

Страница 4: ...when setting the volume proceed as follows when decreasing the volume setting slowly reach the required setting making sure not to overshoot the mark when increasing the volume setting pass the requi...

Страница 5: ...airtight seal FortheP2andP10models adual positionadapter plastic isrequiredtofitDL10tips longtips orD10tips shorttips The metallic rod of the tip ejector is shaped so that the adapter may be clipped t...

Страница 6: ...waste Subsequent volumes that you pipette will have levels of accuracy and precision within speci cations aspirate Press the push button to the rst stop this corresponds to the set volume of liquid H...

Страница 7: ...nd smoothly to the rst stop Wait for at least a second then press the push button to the second stop to expel any residual liquid from the tip Keep the push button pressed fully down and while removin...

Страница 8: ...on slowly and smoothly never turn the pipette upside down never lay the pipette on its side when there is liquid in the tip If you use the same tip with a higher volume pre rinse the tip For volatile...

Страница 9: ...se 7 accessorIes To make pipetting more comfortable and more secure Gilson has developed several accessories Toavoidthepossibilityofliquidrunningbackinto the pipette store the pipette vertically Toide...

Страница 10: ...ice Center 8 GLP FeatUres The Serial Number is engravedonthebodyof the pipette It provides unique identification of your pipette and the date of manufacture Ex CG 12326 The Bar Code on the box and the...

Страница 11: ...der Tighten connecting nut Connecting nut is loose Tighten connecting nut Pipette is Tip holder is loose Tighten connecting nut not precise Connecting nut is loose Tighten connecting nut Incorrect ope...

Страница 12: ...setvolumefromabeakerofdistilled water Maintain the pipette in the vertical position and wait for 20 seconds If a water droplet appears at the end of the tip there is a leak If you see no droplet re im...

Страница 13: ...rformance of your pipette following the veri cation procedure available on the Gilson website www gilson com If the pipette needs to be readjusted please contact your local Gilson authorized Service C...

Страница 14: ...seal Clean autoclave or replace the tip holder Reassemblethepipette refertothe gure p14 Tighten the connecting nut turn clockwise Re t the tip ejector see above Color coded push button Connecting nut...

Страница 15: ...semblethepipette refertothe gure p14 Tighten the connecting nut turn clockwise Re t the tip ejector see above changing the o ring The o ring and seal are on the piston they must not be autoclaved if w...

Страница 16: ...p solution is recommended for cleaning PIPETMAN P external Remove the tip ejector Wipe the tip ejector with a soft cloth or lint free tissue impregnated with soap solution Wipe the entire pipette with...

Страница 17: ...e o ring and seal are not autoclavable they may be cleaned or replaced with the one speci ed in Chapter 14 Spare Parts Clean the parts to be autoclaved especially the tip holder Put the parts in an au...

Страница 18: ...y seal and o ring 13 sPecIFIcatIons PIPETMANPisahighqualitypipettethatoffersexcel lent accuracy and precision The gures given in the GilsonMaximumPermissibleErrors table page19 were obtained using Gil...

Страница 19: ...0 06 0 20 0 10 Min 20 0 35 0 10 0 80 0 30 P100 50 0 40 0 12 0 80 0 30 f123615 Max 100 0 80 0 15 0 80 0 30 Min 50 0 50 0 20 1 60 0 60 P200 100 0 80 0 25 1 60 0 60 f123601 Max 200 1 60 0 30 1 60 0 60 M...

Страница 20: ...Kit 1st level includes 3 piston seals 3 o rings 1 tip holder Service Kit 2nd level includes 1 push button 1 connecting nut 1 tip ejector or only for the P2 and P10 models 1 tip ejector and 1 adapter...

Страница 21: ...Kit 2nd level F161974 F161978 C D Seal O ring 5 sets F144865 F144866 P5000 F123603 and P10ml F161201 Description P5000 P10ml C D E Service Kit 1st level F144499 F144503 A Push button assembly F144787...

Страница 22: ...nts of the following European Directives Villiers le Bel September 14th 2005 a el sayed h Ledorze General Manager Quality Manager GILson s a s 19 avenue des Entrepreneurs BP 145 95400 Villiers le Bel...

Страница 23: ...23 notes...