QUICK GUIDE 

|

 

PIPETMAN® L

Dispense

1. 

Place the end of the tip against the inside wall of the 

recipient vessel (at an angle of 10° to 40°).

2. 

Press the push button slowly and smoothly to the first stop.

3. 

Wait for at least a second, then press the push button to 

the second stop to expel any residual liquid from the tip. 

Keep the push button pressed fully and while removing the 

pipette, draw the tip along the inside surface of the vessel.

4. 

Release the push button smoothly. Eject the tip by pressing 
firmly on the tip ejector button.

Eject the Tips

Before you start to pipette, you can adjust the tip ejector button 

according to your preferences.
1. 

Position the tip ejector button by rotating the tip ejector 

button to the most comfortable position: left, right, or middle.

2. 

Activate the tip ejector. You can either push the tip ejector 

button with the tip of the thumb as usual, or with the base of 

your thumb for more comfort. 

Pre-Wet the Tips

Pre-wetting the tips before pipetting helps prepare the tips for 

the best pipetting performance. Ideally, the pre-wet includes 

both immersing the tip in the liquid and performing one pipetting 

step. Pre-wetting the tips helps ensure that volumes that you 

pipette will achieve accuracy and precision within specifications.

Aspirate

1. 

Press the push button to the 

first stop

 (this corresponds 

to the set volume of liquid).

2. 

Hold the pipette vertically and immerse the tip in the 

liquid (see immersion depth table below).

3. 

Release the push button slowly and smoothly to top 

position to aspirate the set volume of liquid.

4. 

Wait one second (time depends on model, for more 

details, please refer to PIPETMAN L User’s Guide 

LT801575), and then withdraw the pipette tip from the 

liquid. You may wipe any droplets away from the outside 

of the tip using a medical wipe; however, if you do so, 

take care to avoid touching the tip’s orifice. 

Dispense

First stop

Second stop

Rest position

Prepare

Aspirate

Purge

Home

Figure 13

Pipetting motion - Aspirate and dispense

Содержание PIPETMAN L

Страница 1: ...Quick Guide EN A U TOCLAVABL E FULLY VARIABLE VOLUME MODELS...

Страница 2: ...Certificate of Conformity including barcode sticker Tip ejector adapter equipped on P2L P10L Bag of ten filters only for P5000L and P10mLL Multichannel Models PIPETMAN L Quick Guide Safety bag Certifi...

Страница 3: ...ngle Channel Models 2 5 1 2 3 4 1 3 3 2 1 4 4 5 7 8 8 11 11 11 7 8 10 9 6 5 7 1 Color coded push button 2 Thumbwheel 3 Tip ejector button 4 Handle 5 Connecting nut 6 Ejector clip 7 Tip holder s 8 Tip...

Страница 4: ...0 TIPS 5 0 050 0 025 1 0 0 5 0 125 0 075 2 5 1 5 F10L P N FA10020 WITH D10 AND DL10 TIPS 10 0 06 0 03 0 6 0 3 0 12 0 08 1 2 0 8 F20L P N FA10021 WITH D200 TIPS 20 0 10 0 05 0 5 0 25 0 20 0 10 1 0 0 5...

Страница 5: ...0 3 0 2 0 2 0 2 0 1 0 1 0 1 10 0 2 0 1 0 5 0 1 0 0 5 P100L P N FA10004M AND FA10004P WITH D200 TIPS 10 50 100 0 35 0 40 0 80 0 10 0 12 0 15 3 5 0 8 0 8 1 0 0 24 0 15 0 8 0 8 0 8 0 3 0 3 0 3 8 0 1 6 0...

Страница 6: ...20 0 20 20 0 4 0 2 0 10 0 2 0 1 0 P8X200L P N FA10011 AND P12X200L P N FA10012 WITH D200 TIPS 20 100 200 0 50 1 00 2 00 0 25 0 40 0 50 2 5 1 0 1 0 1 25 0 4 0 25 3 2 3 2 3 2 1 2 1 2 1 2 16 0 3 2 1 6 6...

Страница 7: ...key in either direction equals 0 05 L F1L to F2L 0 2 L F5L to F10L 0 5 L F20L to F25L 2 0 L F50L to F100L 5 1 L F200L to F250L 20 L F300L to F1000L 102 L F5000L The cover is graduated in tenths of th...

Страница 8: ...MAN DIAMOND Tip Compatibility for multichannel models P8x10L P12x10L D10 DL10 DF10ST DFL10ST P8x20L P12x20L DL10 D200 DFL10ST DF30ST P8x200L P12x200L D200 D300 DF200ST DF300ST P8x200LVR P12x200LVR D20...

Страница 9: ...11 To fit a tip ejector extension 1 Slide the extension over the tip holder 2 Push the extension firmly onto the end of the tip ejector until it clicks into place To remove a tip ejector extension 1...

Страница 10: ...of your thumb for more comfort Pre Wet the Tips Pre wetting the tips before pipetting helps prepare the tips for the best pipetting performance Ideally the pre wet includes both immersing the tip in...

Страница 11: ...ecommendations given in the user s guide including but not limited to pipettes which have been subjected to physical damage improper handling or spillage or exposure to any corrosive environment This...

Страница 12: ...gilson com contactus LT801634 B 2021 Gilson Inc All rights reserved...

Отзывы: