Gilson Company, Inc. 

Performer III 3in Sieve Shaker: SS-3

Page 5

ASSEMBLING THE SIEVE STACK

Assemble the sieve stack by placing the desired sieves on the 

collection pan. Start with the finest mesh sieves, placing pro-

gressively coarser sieves on top (see Figures 4 and 5). A maxi-

mum of seven metal-framed sieves or fourteen acrylic sieves 

may be stacked on top of the pan. Use of adapters and spacers 

will limit the number of sieves that can be used. The sieve cover 

must be placed over the top sieve before placing the stack as-

sembly in the Performer lll. The larger diameter side of the cover 

is for acrylic sieves, and the smaller diameter for metal sieves. 

Check for the best fit and install.

Hold in the thumb buttons to adjust the clamping knobs on 

the upright rods to the height that will allow the sieve stack to 

slide into place. Pushing in on the integral buttons allows the 

threaded clamp knobs to slide freely. Position the stack using 

the rubber bumpers on the back of the sieve base and tapper 

clamping assembly, then tighten the knobs. To remove the sieve 

stack, hold the thumb buttons in and slide the top up slightly. 

The clamping assembly will remain in this position, ready for the 

next sieve stack. This is a useful feature when using the same size 

stack repetitively.

OPERATING INSTRUCTIONS

To begin testing, set the Mode Switch to MANUAL (see Figure 6). 

Adjust the Amplitude Control for optimum performance with the 

material being tested. Due to varying sample characteristics, the 

user should observe the minimum vibration level and time period 

necessary to sift particles without breaking them down. It is possible 

for excessive vibratory action to degrade the sample particles by 

abrasion.

To change the time setting on the Timer, press the <UP ARROW> 

key or the <DOWN ARROW> key (see Figure 5). The first digit 

on the right hand side should start to flash in half second inter-

vals. Press the <UP ARROW> key or the <DOWN ARROW> key to 

adjust the digit to desired value. To enter the current digit and 

move to the next digit press the <START> key. Once the last digit 

on the left is entered the timer is ready to start.

To get into the adjust mode state, press and hold both the <UP 

& DOWN ARROW> keys at the same time until the current mode 

is displayed. When in this state the display will show the current 

mode letters, H for hours, M for minutes, and S for seconds. Once 

the mode letters are being displayed, press the <UP> key, or the 

<DOWN> key, to change between modes. Press the <START> 

key to accept new mode.

Mode Letter:

A-MMSS B-HHMM          C-SSSS 

        D-MMMM  

Metal Sieves 

Acrylic Spacers

(7 Total)

Sieve Cover

Collection Pan

Figure 3

Figure 4

Acrylic Sieves 

(14 Sieves Max.)

Sieve Cover

Collection Pan

Acrylic Adapter

Содержание Performer III SS-3

Страница 1: ...ve Shaker SS 3 OPERATING MANUAL Rev 01 2019 PHONE 800 444 1508 P O Box 200 Lewis Center Ohio 43035 0200 740 548 7298 E mail customerservice gilsonco com Product Web Page www globalgilson com FAX 800 2...

Страница 2: ...ternative power source is required ALWAYS make sure the available power supply matches the device requirements Motors are NOT explosion proof 2 ALWAYS check electrical wiring for loose connections and...

Страница 3: ...aker SS 3 Page 3 TABLE OF CONTENTS Page Safety Instructions 2 Table of Contents 3 1 0 Introduction 4 2 0 Unpacking Set Up 4 3 0 Assembling the Sieve Stack 5 4 0 Operating Instructions 5 5 0 Specificat...

Страница 4: ...gital timer control ler and graduated vibration control knob allow the Performer lll to give reliable repeatable results The optional GAA 88 Acrylic Spacer is useful for observing sample action during...

Страница 5: ...using the same size stack repetitively OPERATING INSTRUCTIONS To begin testing set the Mode Switch to MANUAL see Figure 6 Adjust the Amplitude Control for optimum performance with the material being t...

Страница 6: ...nue The Setting and the Mode values are saved automatically and restored on power up SPECIFICATIONS Overall Size 8x11x22in 203x279x559mm Vibratory Frequency 3 600vpm Tapping Frequency 60 Taps min Desi...

Страница 7: ...el Figure 7 Performer III Components Accessories Mode Switch Timer Amplitude Control Sieve Base PARTS DIAGRAMS Acrylic Sieve Adapter Sieve Cover 1 4 20 x 3 8in Panhead Screws Sieve Base Tapper Assembl...

Страница 8: ...ompliance to the appropriate ASTM Standard Clear Acrylic Spacers are available if fewer sieves are used in the stack Acrylic Frame Sieves with Stainless Steel Mesh Acrylic Frame Precision Sieves with...

Страница 9: ...15U V3SF 50 V3SF 280U V3SH 18 V3SH 900U V3SH 20 V3SH 800U V3SH 25 V3SH 630U V3SH 30 V3SH 560U V3SH 35 V3SH 450U V3SH 40 V3SH 450U V3SH 45 V3SH 315U V3SH 50 V3SH 280U V3CF 18 V3CF 900U V3CF 20 V3CF 800...

Отзывы: