MyPIPETMAN®

 

|

 

USER’S GUIDE

Specifications

MyPIPETMAN is a high-quality pipette that has excellent accuracy and precision, especially when 

used with PIPETMAN® DIAMOND Tips. The data given in the “Gilson Maximum Permissible Errors” 

table below was obtained using PIPETMAN DIAMOND Tips. These values are guaranteed only when 

genuine PIPETMAN DIAMOND Tips are used.
Each pipette is inspected and validated by qualified technicians in accordance with the Gilson 

quality system. Gilson declares that its manufactured pipettes comply with the requirements of the 

ISO 8655 standard by type testing.
The adjustment is carried out under strictly defined and monitored conditions (ISO 8655-6).

MYPIPETMAN® SINGLE CHANNEL MODELS 

Model

PIPETMAN® 

DIAMOND 

Tips

Vol.

(µL)

Maximum Permissible Errors

Gilson

ISO 8655

Systematic 

Error 

(µL) 

Random 

Error 

(µL)

Systematic 

Error 

(%) 

Random 

Error 

(%CV*)

Systematic 

Error 

(µL) 

Random 

Error 

(µL)

Systematic 

Error 

(%) 

Random 

Error 

(%CV*)

MyPIPETMAN  
P2

D10 
DL10

0.2 

0.5 

2

± 0.024 

± 0.025 

± 0.027 

± 0.030

 0.012 

 0.012 

 0.013 

 0.014

± 12.0

± 5.0

± 2.7

± 1.5

 6.0

 2.4

 1.3

 0.7

± 0.08 

± 0.08 

± 0.08 

± 0.08

 0.04 

 0.04 

 0.04 

 0.04

± 40.0

± 16.0

± 8.0

± 4.0

 20.0

 8.0

 4.0

 2.0

MyPIPETMAN 
P10

D10 
DL10

10

± 0.025 

± 0.075 

± 0.100

 0.012 

 0.030 

 0.040

± 2.5 

± 1.5 

± 1.0

 1.2 

 0.6 

 0.4

± 0.12 

± 0.12 

± 0.12

 0.08 

 0.08 

 0.08

± 12.0 

± 2.4 

± 1.2

 8.0 

 1.6 

 0.8

MyPIPETMAN 
P20

D200

10 

20

± 0.10 

± 0.10 

± 0.20

 0.030 

 0.050 

 0.060

± 5.0

± 1.0

± 1.0

 1.5

 0.5

 0.3

± 0.2 

± 0.2 

± 0.2

 0.1 

 0.1 

 0.1

± 10.0

± 2.0

± 1.0

 5.0

 1.0

 0.5

MyPIPETMAN 
P100

D200

10 

50 

100

± 0.35 

± 0.40 

± 0.80

 0.10 

 0.12 

 0.15

± 3.5

± 0.8

± 0.8

 1.0

 0.24

 0.15

± 0.8 

± 0.8 

± 0.8

 0.3 

 0.3 

 0.3

± 8.0

± 1.6

± 0.8

 3.0

 0.6

 0.3

MyPIPETMAN 
P200

D200

20 

100 

200

± 0.50 

± 0.80 

± 1.60

 0.20 

 0.25 

 0.30

± 2.5

± 0.8

± 0.8

 1.0

 0.25

 0.15

± 1.6 

± 1.6 

± 1.6

 0.6 

 0.6 

 0.6

± 8.0

± 1.6

± 0.8

 3.0

 0.6

 0.3

MyPIPETMAN 
P1000

D1000

100 

500 

1000

± 3.0 

± 4.0 

± 8.0

 0.6 

 1.0 

 1.5

± 3.0

± 0.8

± 0.8

 0.6

 0.2

 0.15

± 8.0 

± 8.0 

± 8.0

 3.0 

 3.0 

 3.0

± 8.0

± 1.6

± 0.8

 3.0

 0.6

 0.3

MyPIPETMAN 
P5000

D5000

500 

2500 

5000

± 12 

± 15 

± 30

 3 

 5 

 8

± 2.4

± 0.6

± 0.6

 0.6

 0.2

 0.16

± 40 

± 40 

± 40

 15.0 

 15.0 

 15.0

± 8.0

± 1.6

± 0.8

 3.0

 0.6

 0.3

MyPIPETMAN 
P10mL

D10mL

1000 

5000 

10000

± 30 

± 40 

± 60

 6 

 10 

 16

± 3.0

± 0.8

± 0.6

 0.6

 0.2

 0.16

± 60 

± 60 

± 60

 30.0 

 30.0 

 30.0

± 6.0

± 1.2

± 0.6

 3.0

 0.6

 0.3

MYPIPETMAN® MULTICHANNEL MODELS

Model

PIPETMAN® 
DIAMOND 
Tips

Maximum Permissible Errors

 Vol.
(µL)

Gilson

ISO 8655

Systematic 

error 

(µL)

Random 

error 

(µL)

Systematic 

error 

(%)

Random 

error 

(%CV*)

Systematic 

error 

(µL)

Random 

error 

(µL)

Systematic 

error 

(%)

Random 

error 

(%CV*)

MyPIPETMAN 
P8x10

D10  
DL10

0.5

1

5

10

± 0.08

± 0.08

± 0.20

± 0.20

 0.04

 0.05

 0.10

 0.10

± 16.0

± 8.0

± 4.0

± 2.0

 8.0

 5.0

 2.0

 1.0

± 0.24

± 0.24

± 0.24

± 0.24

 0.16

 0.16

 0.16

 0.16

± 48.0

± 24.0

± 4.8

± 2.4

 32.0

 16.0

 3.2

 1.6

MyPIPETMAN 
P12x10
MyPIPETMAN 
P8x20

DL10 
D200

2

10

20

± 0.10

± 0.20

± 0.40

 0.08

 0.10 

 0.15

± 5.0

± 2.0

± 2.0

 4.0

 1.0

 0.75

± 0.4

± 0.4

± 0.4

 0.2

 0.2

 0.2

± 20.0

± 4.0

± 2.0

 10.0

 2.0

 1.0

MyPIPETMAN 
P12x20
MyPIPETMAN 
P8x200

D200

20

100

200

± 0.50

± 1.00

± 2.00

 0.25

 0.40

 0.50

± 2.5

± 1.0

± 1.0

 1.25

 0.4

 0.25

± 3.2

± 3.2

± 3.2

 1.2

 1.2

 1.2

± 16.0

± 3.2

± 1.6

 6.0

 1.2

 0.6

MyPIPETMAN 
P12x200
MyPIPETMAN 
P8x300

D300

20

30

150

300

± 1.00

± 1.00

± 1.50

± 3.00

 0.35

 0.35

 0.60

 1.00

± 5.0

± 3.3

± 1.0

± 1.0

 1.75

 1.17

 0.4

 0.33

± 8.0

± 8.0

± 8.0

± 8.0

 3.0

 3.0

 3.0

 3.0

± 40.0

± 26.67

± 5.33

± 2.67

 15.0

 10.0

 2.0

 1.0

MyPIPETMAN 
P12x300

*CV means Coefficient of Variation

NOTE

The data given in this table conform to the ISO 8655-2 standard. With a precise pipetting technique, 

the P2 model can be used to aspirate volumes as low as 0.1 μL and the P10 model as low as 0.5 μL.

Содержание MyPipetman

Страница 1: ...User s Guide EN A U TOCLAVABL E FULLY...

Страница 2: ...r Good Pipetting 10 ACCESSORIES 11 MyPIPETMAN Accessories for Extra Personalization 11 TROUBLESHOOTING 14 Troubleshooting Table 14 Leak Test 15 CLEANING AND DECONTAMINATION 15 Cleaning for Single Chan...

Страница 3: ...tte Colors Special Text Printing Special Logo Printing Ordering Information Minimum Order Quantity Available Countries 1 Worldwide Possible on Request Subject to Specific Conditions 150 Units Mix of D...

Страница 4: ...and MyPIPETMAN Enterprise serial number type A A 50001 MyPIPETMAN serial number type MY 524 A A Figure 2 Serial number identification example Year Month Production number Model Random Number Month Ye...

Страница 5: ...6 0 3 MyPIPETMAN P200 D200 20 100 200 0 50 0 80 1 60 0 20 0 25 0 30 2 5 0 8 0 8 1 0 0 25 0 15 1 6 1 6 1 6 0 6 0 6 0 6 8 0 1 6 0 8 3 0 0 6 0 3 MyPIPETMAN P1000 D1000 100 500 1000 3 0 4 0 8 0 0 6 1 0 1...

Страница 6: ...00 L 0 2 L MyPIPETMAN P8x300 L 0 2 L MyPIPETMAN P12x300 L 0 2 L 1 5 P2 1 25 L 2 0 7 5 P10 7 5 L 1 2 5 P20 12 5 L P100 0 7 5 75 L P200 1 2 5 125 L P1000 0 7 5 0 75 mL P5000 1 2 5 1 25 mL P10mL 0 7 5 7...

Страница 7: ...r extension refer to Figure 7 1 Slide the extension over the tip holder 2 Push the extension firmly onto the end of the tip ejector until it clicks into place To remove a tip ejector extension 1 Gentl...

Страница 8: ...accuracy and precision by creating an efficient barrier to aerosols or corrosive vapors such as acids which could damage mechanical parts Insert the filter directly in the pipette tip holder Change th...

Страница 9: ...wall of the recipient vessel at an angle of 10 to 40 2 Press the push button slowly and smoothly to the first stop 3 Wait for at least a second then press the push button to the second stop to expel...

Страница 10: ...the tip For volatile solvents saturate the air cushion in your pipette by aspirating and dispensing the solvent repeatedly before aspirating the sample When the liquid temperature is different from t...

Страница 11: ...nap cap and screw cap microtubes Description Part Number JIMMY Microtube Opener set of 3 F144983 For P5000 and P10mL models disposable filters maintain the pipette s original accuracy and precision by...

Страница 12: ...P200 MyPIPETMAN FP070142 Metal Ejector Dark Blue P1000 MyPIPETMAN FP070151 Metal Ejector White P1000 MyPIPETMAN FP070152 PLASTIC TIP EJECTORS Model Part Number Plastic Ejector Dark Blue P2 P10 MYPIPET...

Страница 13: ...Holder Dark Blue P1000 MyPIPETMAN FP070281 Tip Hholder White P1000 MyPIPETMAN FP070282 Tip Holder Dark Blue P5000 MyPIPETMAN FP070291 Tip Holder White P5000 MyPIPETMAN FP070292 Tip Holder Dark Blue P...

Страница 14: ...place the tip holder S Connecting nut is loose Tighten connecting nut S M Damaged or corroded piston Return pipette to supplier S Improper repair or assembly Refer to MAINTENANCE S Pipetted volume is...

Страница 15: ...pipette to the maximum volume given in the specifications and pre wet the tips 3 Aspirate the set volume from a reagent reservoir of distilled water 4 Maintain the pipette in the vertical position and...

Страница 16: ...or wipe them with a soft cloth or lint free tissue soaked with the cleaning solution 3 Rinse the components with distilled water 4 Wipe the entire pipette with a soft cloth or lint free tissue soaked...

Страница 17: ...e being immersed in decontamination solution 1 Disassemble the pipette refer to MAINTENANCE 2 Immerse tip ejector tip holder and connecting nut in the decontamination solution 3 Disassemble and degrea...

Страница 18: ...p ejector along the tip holder 3 Clip the tip ejector on the body of the pipette Changing the Tip Holder for Single Channel 1 Push the ejection button 2 Turn the tip ejector clockwise 3 Slide and remo...

Страница 19: ...p holder from the upper part refer to Figure 11 1 Remove the piston assembly O ring and the seals refer to Changing the Tip Ejector for Single Channel 2 Leave the piston exposed clean it with isopropa...

Страница 20: ...MAN FP070105 Window 3digit Green MyPIPETMAN FP070106 Lubricant Tube for PIPETMAN 1g 5440011070 NOTICE For pipette service and maintenance or for extra spare parts please contact your local Gilson Serv...

Страница 21: ...ilson s sole liability under this warranty shall be limited to at Gilson s sole option repair or replacement of any defective components of pipettes or refund of the purchase price paid for such pipet...

Страница 22: ...ilson Inc All rights reserved Specifications subject to change without notification errors excepted www gilson com contactus LT801622 D 2022 Gilson Inc All rights reserved Specifications subject to ch...

Отзывы: