6

MACROMAN

TM

 

|

 

USER’S GUIDE

INTRODUCTION

 

MACROMAN

TM

INTRO

D

U

C

TI

O

N

Component Description

COMPONENT

MATERIAL / PROPERTIES

Housing

UV-resistant resin

Nosepiece

UV-resistant resin

Thumbwheel

UV-resistant resin

Side stand

UV-resistant resin

Pipette Adaptor

Silicone

Filter

PTFE, standard or sterile, hydrophobic membrane filter, 

0.22 µm or 0.45 µm

Wall mount and Desk Stand

ABS (Acrylonitrile butadiene styrene)

 

Technical Information

Technical Data

Flow rate

<3 seconds with 0.45 µm filter for 25 mL 

<6 seconds with 0.45 µm filter for 50 mL

Dimensions

3.81  x 14.98 x 14.98 cm (1.5”W x 5.9”D x 5.9”H) 

Weight

MACROMAN: 205g

Power supply

Input voltage: 100–240 VAC, 50/60 Hz, 0.3A  

Output voltage: 5 VDC, 500 mA Center positive DC connector

Unit: Input voltage

5 VDC, 500 mA Center positive DC connector

Battery

Lithium rechargeable battery 3.7 V/ 2800 mAh

Battery charging time

 6 hours

Battery discharge time

> 10 hours

Equipment Disposal

This equipment must not be disposed of with unsorted municipal 

waste. Instead, it is your responsibility to correctly dispose of 

your waste equipment by handing it over to an authorized facility 

for separate collection and recycling (do not dispose the device in a fire; 

do not modify the Lithium-Ion battery in any way; discharge the battery 

before disposal; dispose of the MACROMAN and the battery separately in 

accordance with the laws and regulations in your area governing disposal 

of devices containing Lithium-Ion batteries). It is also your responsibility to 

decontaminate the equipment in case of biological, chemical,  

and/or radiological contamination so as to protect from health hazards the 

persons involved in the disposal and recycling of equipment. 

For more information about where you can drop off your waste equipment 

for recycling, please contact your local dealer from whom you originally 

purchased the product or your local council. By doing so, you will help 

conserve natural resources and you will ensure that your waste equipment 

is recycled in a manner that protects human health and the environment. 

Thank you.

O

PE
R

A

TIO

N

WARNING

Ensure that you are using an original Gilson main power supply  

(100–240 VAC, 50/60 Hz) — part number F107203. 

The power supply must not to be used for any other electrical equipment 

except MACROMAN.

Chapter 2

OPERATION

Operating Conditions

SPECIFICATION

 CONDITION

Operating temperature

+10°C to +40°C

Relative humidity

10% to 90% (no condensation)

Power supply

Fluctuations not to exceed 10%

Charging MACROMAN

1.  Before using MACROMAN for the first time, insert 

the proper adaptor attachment into the power 

supply (Figure 3). Turn the plug clockwise to lock.

2.  Release by pushing down the button and turn 

counter-clockwise to release the plug.

3.  Plug the power supply into a standard 

electrical source.

4.  Gently insert the power jack into the 

power supply charging socket located in 

the base of MACROMAN (Figure 4).  

 

5.  When MACROMAN is properly connected, 

the battery icon will indicate the charging 

status.

6.  The battery display shows between 1 to 

6 bars in sequence from empty to full. 

Allow the battery to charge for more 

than 6 hours before using MACROMAN 

for the first time. Once MACROMAN is 

completely charged, the battery icon 

shows as full (solid bars). You can also 

use MACROMAN during the charging 

process. 

Using excessive force may 
damage MACROMAN. 

The battery icon will flash when battery is at less than 10% charged.  
The battery is charged at: 

• 

15% after 1 hour

• 

30% after 2 hours

• 

61% after 4 hours

• 

100% after more than 6 hours

Figure 3

Insert the adaptor you need 

into the power supply

Figure 4

Plugging the power supply into 
MACROMAN

TM

MASTER

Figure 5

Battery icon

Содержание macroman

Страница 1: ...User s Guide EN...

Страница 2: ...mation 6 OPERATION 7 Operating Conditions 7 Charging MACROMAN 7 Using MACROMAN 8 MAINTENANCE CLEANING 10 Safety and User Precautions 10 Replacing the Battery 10 Replacing the Filter 11 Cleaning 11 TRO...

Страница 3: ...ease make sure to read the safety and operating instructions before use Also when performing the pipetting function observe all laboratory safety guidelines including the use of protective clothing sa...

Страница 4: ...e you can drop off your waste equipment for recycling please contact your local dealer from whom you originally purchased the product or your local council By doing so you will help conserve natural r...

Страница 5: ...Rotate the pipette adaptor to accommodate different types of pipettes Left position Low Right position High MASTER Figure 9 Adjust speed with the thumbwheel to set the flow rate Figure 7 MACROMANTM o...

Страница 6: ...e 12 MACROMANTM battery A C B Do not force the connector It only fits one way There should be a slight click when the cable connection is inserted properly A C B It is imperative to dispose of a used...

Страница 7: ...ng Table SYMPTOM POSSIBLE CAUSE ACTION The serological pipette falls out the holding force of the pipette is too small or tilts to the side too much The pipette adaptor is dirty or wet Take out the pi...

Страница 8: ...or replacement will be provided by your local Gilson service center Freight shipping the unit to Gilson will be the responsibility of the purchaser There are no other warranties or guarantees implied...

Страница 9: ...01 2018 Gilson Inc All rights reserved Specifications subject to change without notification errors excepted Reproduction adaptation or translation without prior written permission is prohibited exce...

Отзывы: