Gilson HM-380R Скачать руководство пользователя страница 3

3

Stop Switch

  –  Momentary  push  button  to  stop  the  

motor.

Direction Switch

 – Momentary (spring loaded) switch. 

After setting the desired shear rate, moving the switch 

in  the  desired  direction  will  initiate  movement.  This 

switch will then become inactive until the stop button is 

pressed, or a limit switch is activated. The left position 

will move the water chamber to the left. It will continue 

until  the  water  chamber  activates  the  limit  switch. At 

this  time,  the  limit  indicator  light  will  be  lit.  The  right 

switch position will move the shear box to the right until 

the home limit switch is made. See above note…

Home & Limit Indicators

 – Used to indicate the limit 

of travel of the shear box.

Fuse

 – 4 amp SLO BLO fuse located in the cabinet 

rear.
 

OPERATING INSTRUCTIONS

Read all operating instructions before performing a test 

and refer to ASTM D3080 or AASHTO T 236 for com-

plete test procedures and calculations.

Set Up

1.  Place the shear box into the water chamber and se-

cure the load cell shaft to the top of the shear box.

2.  Tighten the two shear ring clamping knobs located 

on the left side of the water chamber.

3.  Install the counterbalance, Model HMA-735, if nec-

essary. 

4.  Tare the load cell by pressing the “Tare” button on 

the digital readout.  

5.  Adjust the top cross arm and dial gauge so there is 

1/8in gap between the cross arm and loading ball. 

Align the cross arm pin so it is above the loading 

ball.  Make sure the cross arm is level.

6.  Adjust the dial gauge by raising or lowering the in-

dicator rod and holder so the tip of the dial gauge is 

resting on the cross arm pin.   

Consolidation

The weight of the consolidation load arm with two up-

right rods lower cross arm and lower weight platform is 

11.9lbs.  The lever arm on the right side of the device 

has been preset to a 10:1 ratio, select a consolida-

tion load and apply the requires weights to the weight 

hanger.  The consolidation result is read on the top dial 

indicator.  
Refer to ASTM D3080 for consolidation calculations.

Direct Shear

1.  After consolidation phase is complete, add a small 

seating shear load by adjusting the two knurled nuts 

on the load cell support.

2.  Adjust or zero the horizontal dial indicator.
3.  Remove the two holding screws on the shear box 

to separate the top and bottom sections.  Adjust the 

four  adjusting  screws  on  the  shear  box  to  set  the 

gap between the top and bottom of the shear box.  

One full turn of these adjusting screws will give a 

0.031in (0.78mm) gap.  

4.  Set the shear/strain  rate with  the thumbwheels  on 

the left side of the machine.  Each number is set by 

pushing the thumbwheel button, the rate is in inches 

per minute (0001 = 0.0001 in/min).

5.  To begin shear, switch the DIRECTION toggle switch 

to  the  left  towards  the  LIMIT  light.  Record  shear 

readings at the prescribed intervals. 

6.  The water chamber with shear box will continue to 

move at the set rate until the limit switch is activated 

or the STOP button is pressed.  

7.  At the test completion, switch the DIRECTION tog-

gle to the right towards the HOME light. Press the 

stop button when Zero load is reached on the load 

cell readout.  

8.   Remove the vertical load by removing the weights 

from the Weight Hanger

9.   The Shear Box can now be removed from the water 

chamber and the sample can be inspected. 

 

MAINTENANCE 

Special care should be taken to see that the four roll-

er bearings between the water chamber and the base 

slide track are properly cleaned and contain a slight 

amount of oil during and after use.  It is recommended 

to cover the Direct Shear machine when not in use to 

prevent dust buildup.
Empty the water chamber after each test to prevent soil 

and sand buildup inside the chamber.  The water can be 

drained from the port, or the chamber can be removed 

from the Direct Shear Machine for thorough cleaning.   

ACCESSORIES

HMA-739 

Shear Box Counterbalance

HMA-610  

Direct/Residual Shear Data Acquisition 

Software

 