PIPETMAN® M CONNECTED 

|

 

USER’S GUIDE

PIPETTING MODES

 

PIPETMAN® M CONNECTED

4.  Fit a tip suitable for the model 

of PIPETMAN M Connected 

that you are using (preferably 

use PIPETMAN DIAMOND Tips 

for assurance of accurate and 

precise results; refer to Chapter 

PiPETMAN diAMONd TiPS

 on 

page 15).

5.  Press the push button to aspirate 

total volume. The volume 

aspirated will be a little bit more 

than required (extra volume). The 

extra volume is necessary to ensure 

equal operating conditions for each 

dispensed aliquot.

dispense

Press the push button. 

diSCArd

 appears on the screen. Discard a part of the extra 

volume and continue dispensing the aliquots. PIPETMAN M Connected distributes the 

aliquot volume each time you click the push button. The number of aliquots left to 

dispense is displayed on the screen. A warning beep indicates the last aliquot.

Purge

Press the push button. 

PUrGE 

appears on the screen. Click again to dispose and then 

purge the rest of the extra volume and reset the piston. Keep your thumb pressing the 

push button until the end of the purge—the piston reset takes place after you release 

the push button again. This feature enables you to remove the tip from the liquid, 

without aspirating anything

.

Eject Tip

 

Press the ejection button.

MIX Mode

This a mixing phase, composed of repeatedly aspirating and dispensing.

1. 

Press the menu button to access menu, where you can choose between different 

pipette modes. Select 

“MiX” Mode 

by turning the push button, and then to confirm 

your selection, click on the push button.

Fit a tip suitable for the model of PIPETMAN M Connected that you are using 

(preferably use PIPETMAN Tips for assurance of accurate and precise results, refer to 

Chapter 9 

PiPETMAN diAMONd TiPS

 on page 15).

1. 

Press the push button to aspirate the selected volume.

2.  To dispense: click (press and release) the push button. 

Three things will happen: sample dispense, automatic 

purge, and piston reset to zero. Or, keep your thumb 

pressing the push button until the end of the dispense 

cycle and the piston reset takes place after releasing the 

push button again. This feature enables you to remove 

the tip from the liquid, without aspirating anything. 

3.  Eject the tip by pressing the ejection button.

Your PIPETMAN M Connected is ready for the next pipetting  

cycle while maintaining the volume and the last settings.

Continue pipetting as you would with your mechanical PiPETMAN; it’s as simple as 

that to use your new, electronic pipette!

REPETITIVE Mode

The 

rEPETiTiVE Mode 

allows you to distribute the same volume repeatedly in a 

predefined number (N) of equal aliquots. You supply the aliquot volume 

(AV) and the pipette automatically calculates the number of aliquots 

possible from the nominal (maximum) volume (NV) of the pipette, as 

follows: 

N = NV/AV 

(e.g., 120 μL x 10 for P1200M). 

You also can decrease the number of repetitions (e.g., 120 μL x 10-(n 

1) for P1200M).

NOTICE

Always fit a tip 

before using any 

pipette.

Figure 3

REPETITIVE Mode

NOTE

The MiX Mode can be used to prepare a PCr master mix, enzyme restriction 

mix, protein solution, oligonucleotide dilution and to mix samples with 

gel loading buffer and so on… you can mix two different solutions or 

homogenize one solution. if you work with higher viscosity than water (e.g., 

restriction enzyme), you may change the aspiration speed (refer to Chapter 8 

PiPETTiNG SPEEd CONTrOL 

on page 15). if you work with genomic dNA, then 

you should always pipette very carefully to avoid shearing and nicking.

NOTICE

Always fit a tip before using any pipette.

START button
Optional STOP

0

V

Purge

Figure 2

PIPET Mode

NOTE

The rEPETiTiVE Mode is ideal for dispensing aliquots. Examples include 

dispensing a PCr master mix into PCr tubes or 96-well, pipetting elution buffer 

for dNA extraction, preparing a target for spectrometric analyses, distributing 

loading buffer into samples, etc. 

1. 

Press the menu button to access the system menu, where you can choose between 

different pipette modes. Select 

rEPETiTiVE Mode

 by turning the push button.  

To confirm your selection, click on the push button.

2.  Turn the push button half way, the screen will display:  

“Click to change volume.”

3.  Press the push button; the volume on the display will start to blink, 

you can now adjust the volume.

Set the Aliquot Volume

1. 

Click the push button, set the volume by turning the push button and click again 

to confirm your settings. 

2.  Set the aliquot number: the maximal number of aliquots has been calculated 

by your PIPETMAN M Connected. The number of aliquots flashes and you can 

decrease this number (N – (n 

1)) by turning the push button. 

3.  Press again to confirm your settings.

Содержание F81040

Страница 1: ...User s Guide EN...

Страница 2: ...D 18 Tip Ejector Extension for Use with P10M 18 Personal Label Name and or Application Tag 19 GOOD LABORATORY PRACTICE FEATURES 19 General 19 Cycle Counters 19 Alarms 20 POWER MANAGEMENT 20 Battery Ch...

Страница 3: ...fatigue and increase pipetting efficiency PIPETMAN M Connected is a unique pipette with only two buttons to operate the pipette and reach all menu options PIPETMAN M Connected features the classic PIP...

Страница 4: ...9 10 11 12 13 14 15 16 17 18 19 Upper part Push button and volume adjustment knob Menu button direct access to all menu functions Tip ejector button Display Elastomer plug to cover the battery chargi...

Страница 5: ...the tip by pressing the ejection button Your PIPETMAN M Connected is ready for the next pipetting cycle while maintaining the volume and the last settings Continue pipetting as you would with your mec...

Страница 6: ...g protocol first change the pipetting speed of your pipette refer to Chapter 8 PIPETTING SPEED CONTROL on page 15 Aspiration can be done by using the standard speed by default speed 6 Dispensing shoul...

Страница 7: ...nables personalized pipetting protocol creation on a PC or tablet for transfer to PIPETMAN M Connected through PIPETMAN M Connected Software an easy to use software To use this mode on a PC the softwa...

Страница 8: ...pending on your application You can change aspiration speed and dispensing speed independently from very slow to very fast speed 1 speed 6 Your speed selection is memorized for each pipetting mode unt...

Страница 9: ...4 5 P10mLM 5 7 4 5 P8x10M P12x10M 1 1 P8x20M P12x20M 2 3 1 P8x100M P12x100M 2 4 1 P8x200M P12x200M 2 3 1 P8x300M P12x300M 2 3 1 P8x1200M P12x1200M 2 4 2 3 Table 1 Immersion depth and wait time Mold an...

Страница 10: ...surface of the liquid refer to Immersion Depth and Wait Time Table 1 on page 17 Change the tip before aspirating a different liquid sample or reagent Change the tip if a droplet remains at the end of...

Страница 11: ...thium Ion battery Operation on the battery is done at user s risk only Opening the pipette voids the warranty Alarms Low battery warning Service is overdue warning refer to Chapter 15 MAINTENANCE on p...

Страница 12: ...mber of readjustments carried out by this pipette NOTE Changes of service settings will reset all the service info to zero Chapter 14 CONFIGURATION Enter the menu using the menu button You will find n...

Страница 13: ...tion Enter the sub menu in the same way and select the ADJUSTMENT menu The adjustment menu allows you to access the following items Standard readjustment Reset settings Standard Readjustment This menu...

Страница 14: ...Chapter 21 REPLACEMENT PARTS on page 34 To access the O ring remove the tip ejector and unscrew the lower part of the tip holder You should now be able to remove the O ring from the piston Sometimes t...

Страница 15: ...er 16 CLEANING AND DECONTAMINATION PIPETMAN M Connected is designed so that the parts normally in contact with contaminants can easily be cleaned and decontaminated As part of your quality system you...

Страница 16: ...nents with distilled or sterile water 4 Leave the parts to dry by evaporation or wipe them with a clean lint free tissue or soft cloth 5 Lubricate the piston and reassemble the pipette according to th...

Страница 17: ...e used with the recommended PIPETMAN DIAMOND Tips Under these conditions Gilson volumetric specifications in standard pipetting PIPET mode are guaranteed with a performance exceeding ISO 8655 2 recomm...

Страница 18: ...22 F807022 F807022 F807022 F807022 F807022 F807022 Battery window F807005 F807005 F807005 F807005 F807005 F807005 F807005 F807005 Ejector window F807006 F807006 F807006 F807006 F807006 F807006 F807006...

Страница 19: ...ture and pressure for distilled water TEMPERATURE AIR PRESSURE HPA C 800 853 907 960 1013 1067 15 1 0018 1 0018 1 0019 1 0019 1 0020 1 0020 15 5 1 0018 1 0019 1 0019 1 0020 1 0020 1 0021 16 1 0019 1 0...

Страница 20: ...R IMPLIED INCLUDING ANY WARRANTIES OF MERCHANTABILITY AND ANY WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE UNDER NO CIRCUMSTANCES SHALL GILSON BE LIABLE FOR ANY CONSEQUENTIAL PUNITIVE INDIRECT OR IN...

Страница 21: ...gilson com contactus LT801594 G 2021 Gilson Inc All rights reserved...

Отзывы: