4223
Lightweight Liquid Level Sensor
Gill Sensors & Controls Limited
4.3 FITTING THE MOUNTING ADAPTOR
Fitting the 19 x 1.5 mm fluorosilicone O-ring seal
Prior to installation of the sensor in the tank an fluorosilicone O-ring seal and/or fluorosilicone
panel gasket must be fitted to ensure the tank is sealed. All sensors must be fitted with the 19
x 1.5 mm fluorosilicone O-ring seal as shown in the diagram below.
Fitting the sensor adaptor
The sensor must then be fitted with the adaptor (1, 2, 3, 5 hole or threaded adaptor) using the
three supplied M3 x 6 mm CSK screws as shown in the diagram below. The adaptor has 3
possible orientations for 1, 2, 3 hole and threaded adaptors - for the correct orientation of the
5 hole adaptor see the diagram on the next page. (Note: Thread lock pre-applied to screws.
Tighten to 1.3 Nm with 2.0 mm Hex Key.)
4223-30-090 lss. 1
10