R3-50, R3-100 and R3A-100 Doc. No 1210-PS-0015 Issue 2
June 2021
___________________________________________________________________________________________
Page 17
8.11 Interactive Mode
Interactive mode is entered from measurement mode, by sending the anemometer the IM
command (see Para 14.2). In this mode the user can communicate with the anemometer, issuing
configuration commands to the anemometer, or requesting information from the anemometer.
8.12 Configuration Commands
Configuration commands (see Para 14.2) allow the user to configure the anemometer as required.
The configuration parameters are held in non-volatile storage, allowing the anemometer to retain its
current configuration when powered down.
8.13 Default Configuration Command Settings
Factory settings are:-
AVERAGE
5
WINDREP
UVW CAL
SOSREP
SONICTEMP K
ABSTEMP
OFF
ANAIP
DDDDDD (DISABLED)
AOPFSD
30
POLARWRAP
360
STRFMT
ASCII UNPAD
ASCTERM
CRLF
ECHO
ON
MSGMODE
CONT
BAUD
9600
CTONE
ON
ALIGN
AXIS
INST
OFF
8.14 Default Output Data String
The Structure of the Factory R3 Default Output Data String is as follows:-
Status Codes, U, V, W, Speed of Sound, Checksum.
01,00,-00.04,+00.00,+00.03,293.94,1E
02,28,-00.04,+00.00,+00.03,293.94,17
03,00,-00.04,-00.02,+00.03,293.94,18
04,00,-00.05,-00.02,+00.04,293.94,19
05,00,-00.04,-00.03,+00.03,293.95,1E
06,01,-00.05,-00.02,+00.04,293.94,1A
8.15 Interrogation Commands
Interrogation commands allow the user to request various information from the anemometer.
E.g.
?, CONFIG, EXIT, HELP, IM, SN, VER and VERIFY.
See Para 14.2.