Set Flow Rate 

1.

 

From the 

Main Menu

, select 

Flow Set

 using the 

NAV

 keys. 

2.

 

Press the 

INC/DEC

 keys to set desired flow rate, then press the 

POWER/ENTER

 key to enter the desired flow rate. 

 
Note:

 Flow range select must be in Lo if the flow is less than 445 cc/min and in 

Hi for rates from 450 to 5100 cc/min. The flow range select is on the right side 
of the case and can be switched between Lo and Hi with a 2mm (5/64") hex 
wrench supplied with the pump. Indication of Lo or Hi selection displays in the 
middle of the lower display status line. 

 

 

Flow Calibration 

1.

 

Connect the pump to an air flow calibrator per the 
calibrator manufacturers’ recommendations.  A 
representative sample media must be connected at 
the inlet to the pump to establish proper load 
conditions.  A calibration panel may be substituted 
for the sample media, set for a pressure drop of 4 
inches of water.  

2.

 

While pump displays the 

Main Menu

 use the 

NAV

 

keys to select 

Calibrate, 

then

 

Press the 

POWER/ENTER

 key. 

3.

 

Display shows the set flow rate and the pump 
begins to run in the calibration mode. 

4.

 

Use the 

INC/DEC

 keys to adjust the flow rate 

displayed on the pump until matching the flow rate 
measured by the calibration device.   

5.

 

Press the 

POWER/ENTER

 key to set calibration. 

6.

 

Press 

ESC

 key to return to 

Main Menu

.  

 

Field Calibration Note: 

Display calibration procedures above make internal pump adjustments and improve the accuracy of the flow display. This 
does not replace field calibration as described by OSHA and NIOSH. Conduct flow verification using a Primary Calibrator 
prior to each field sample. Field calibration procedures are referenced in the 

NIOSH Manual of Analytical Methods

 at 

www.cdc.gov/niosh and the 

OSHA Technical Manual

 at www.osha.gov.

 

Starting the Sample Run 

1.

 

With pump displaying the 

Main Menu

 use the 

NAV

 

keys to select 

Run. 

 

2.

 

Press the 

POWER/ENTER

 key. 

 Note:  Before it enters the Run mode, the pump 
may go into a self-calibration mode for 7-10 
seconds.  During this interval, "Sensor Calibration" 
displays. 

Stopping the Sample Run 

1.

 

From any display, press 

POWER/ENTER

2.

 

The

 Pause/Stop Menu 

will appear in the upper 

left corner of the display.  

3.

 

Select 

Stop

 and press 

POWER/ENTER 

to stop the 

sample. 

 
 

Retrieving Data 

1.

 

From the 

Main Menu

, using the 

NAV

 keys select 

Review

2.

 

Press 

POWER/ENTER

3.

 

Use the 

NAV

 

keys to select among the last sixteen events. 

 
 

Note:  

Total Run Time and Total Volume Sampled do not reset during 

Pause

. However, 

Stop 

will end the sampling event and the 

data will clear at the next sampling event. Data displays until the next event starts. Select 

Review

 to retrieve previous 

sampling data.  See 

Retrieving the Data

 (above).

 

Содержание PN 360-0132-01

Страница 1: ...PN 610 0901 01 R GilAir Plus Datalog PN 610 0901 02 R GilAir Plus STP PN 610 0901 03 R GilAir Plus Datalog w Bluetooth PN 610 0901 05 R GilAir Plus STP w Bluetooth PN 610 0901 06 R 1000 112th Circle N Suite 100 St Petersburg FL 33716 USA 800 451 9444 1 727 530 3602 www Sensidyne com info Sensidyne com REF 360 0135 01 Rev E ...

Страница 2: ...u Operation Guide Power Pump On and Off Pump should be fully charged before use Power Pump On Press and hold the POWER ENTER key until the pump displays the Start up Screen Power Pump Off When pump is not running or in a program pause press and hold POWER ENTER key Continue to hold POWER ENTER until Power down window appears and countdown completes Cautions Intrinsic Safety The pump is intrinsical...

Страница 3: ...d by the calibration device 5 Press the POWER ENTER key to set calibration 6 Press ESC key to return to Main Menu Field Calibration Note Display calibration procedures above make internal pump adjustments and improve the accuracy of the flow display This does not replace field calibration as described by OSHA and NIOSH Conduct flow verification using a Primary Calibrator prior to each field sample...

Страница 4: ...igure GilAir Plus sampling pumps and manage data collected by the pump Gilian CONNECT Mobile available for Android and iOS can connect to a Bluetooth enabled pump during a sample run allowing confirmation that pump is performing as desired Users may Start Pause and Stop a sample review sample history and document the pump s use by using the mobile device s camera to photograph and embed current ti...

Отзывы: