2.3.3
Sensor Zero and Calibration
All sensors must be calibrated and zeroed before making measurements.
Calibration and Zeroing
1.
Connect the sensor to be calibrated from Channel A or B to the Calibrator output
(see Figure 2-14).
2.
Calibrate the sensor(s) within the 5 °C to 35 °C (41 to 95 °F) operating temperature range of
the Calibrator.
3.
Press the
[CAL/ZERO]
hardkey. The meter will automatically zero and calibrate the sensor.
A sensor requires several minutes to fully zero and calibrate. After the zero/calibration routine
is complete, the display will return to the Main menu.
Zeroing Only
1.
If you need to only zero one or both sensors, make sure both sensors are connected to the sensor
inputs and that neither sensor is connected to the Calibrate output. When zeroing a sensor, it
is best to connect the sensor to the device under test exactly as it will be used in measurement,
and deactivate the RF output of that device. Zeroing the sensor in place is the best way to
counteract system noise which could significantly effect low-level measurements.
NOTE:
Connector sensor to the Calibrator connector after the power meter
is turned on. Allow a 15-minute thermal equilibrium period for the sensor
and calibrator connector to minimize temperature differences.
Figure 2-14: Setup for Sensor Calibration
NOTE:
If the calibrator is set to 0.00 dBm and turned on before or
immediately after calibration, it is possible to measure approximately ±0.02
dBm from the calibrator output instead of 0.00 dBm (0.4% difference). This is
normal and within specification due to a combination of tolerances resulting
from calibrator accuracy (±1.2% worst case for one year, over temperature
range of 5 °C to 35 °C) and connection tolerance.
☛
☛
☛
☛
40%
☛
☛
☛
☛
Содержание 8650A Series
Страница 4: ......
Страница 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Страница 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Страница 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Страница 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Страница 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Страница 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Страница 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Страница 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Страница 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...