Manual 31470, Rev. E, April 2001
4.4.3
GPIB Port Check
The following steps confirm that the GPIB port is functional:
1.
Set the 8650A to the desired address (the default is 13; see Section 2.2.2 for instructions on
how to set the GPIB mode and address).
2.
Connect the GPIB controller to the GPIB Port on the rear of the 8650A.
3.
Send the command (if emulating an HP 438):
*IDN? or ?ID
(*IDN? is the IEEE 488.2 Common ID query. When addressed to talk after receiving the com-
mand, the 8650A will output a string that identifies itself as Model 8651A or 8652A.)
4.
Display the response on the controller.
This completes the Specification and Performance Verification Test for the 8650A Universal Power
Meter and its sensors. If the instrument has performed as described in this chapter, it is correctly
calibrated and within specifications.
Содержание 8650A Series
Страница 4: ......
Страница 12: ...Series 8650A Universal Power Meters viii Manual 31470 Rev E April 2001 ...
Страница 16: ...Series 8650A Universal Power Meters xii Manual 31470 Rev E April 2001 ...
Страница 18: ...Series 8650A Universal Power Meters xiv Manual 31470 Rev E April 2001 ...
Страница 20: ...Series 8650A Universal Power Meters xvi Manual 31470 Rev E April 2001 ...
Страница 32: ...Series 8650A Universal Power Meters 1 12 Manual 31470 Rev E April 2001 ...
Страница 198: ...Series 8650A Universal Power Meters 3 124 Manual 31470 Rev E April 2001 ...
Страница 216: ...Series 8650A Universal Power Meters A 8 Manual 31470 Rev E April 2001 ...
Страница 228: ...Series 8650A Universal Power Meters B 12 Manual 31470 Rev E April 2001 ...
Страница 240: ...Series 8650A Universal Power Meters Index 6 Manual 31470 Rev E April 2001 ...