8003 Precision Scalar Analyzer
2-60
Manual 20791, Rev. C, June 2001
[
CW FREQ]
Menu SY1-2:
Directly specifies the frequency of the signal to the sensor. When this softkey is
pressed, the entry area of the display will show the currently defined signal frequency and present a
prompt to change it as desired. Once this softkey has been pressed, the instrument will assume that
all measurements made in the CW mode on the setup sensor are at the specified CW frequency.
The instrument will correct the power readings of the sensor according to the frequency correction
table stored in the sensor’s PROM.
[
CW CAL FACTOR]
Menu SY1-2:
Specifies user-defined sensor frequency response correction factors. When this soft-
key is pressed, menu SY1-4 will be shown allowing the selection of the sensor to which the cal fac-
tor will be applied. Once the selected sensor is specified, the instrument will apply the cal factor to
all CW measurements made with the sensor, and the frequency correction table stored in the sen-
sor’s PROM will not be used when in the CW mode.
[
FROM SOURCE]
Menu SY1-2:
Specifies that the 8003 should automatically determine the frequency of the signal
from the signal source. Pressing this softkey will cause menu SY1-5 to appear wherein either all or
any individual sensors can be selected to determine the source’s CW output frequency as set by the
8003 through the source control keys.
[
CORRCTN OFF]
Menu SY1-2:
Accesses menu SY1-6 wherein the correction factor for either all or any individual
sensors can be turned off.
[
SENSOR OFFSET]
Menu SY1:
Causes menu SY1-7 to appear. This menu either clears or enters offsets for the selected
sensors. When the selected SENSOR OFFSET softkey is pressed, the entry area of the display will
show the current value of the setup sensor’s offset. The offset can then be changed as desired.
The SENSOR OFFSET softkeys of menu SY1-7 will be in light blue if the setup sensor’s offset is
nonzero. If the sensor offset is zero, the softkey label will be in standard menu green since no offset
is being used.
[
PULSE SENSORS]
Menu SY1:
Accesses menu SY1-8 and then SY1-9 wherein channel and operating modes can be
specified when any of the 80340 Series of Triggerable Pulse (Peak Power) Sensors are used with the
8003. The channel of interest is first selected and then either the Free Run, Trig’d, or Delay Trig’d
mode is specified in menu SY1-9. In the Free Run mode, the RF power is sampled automatically and
the sensor does not require a trigger signal. In the Trig’d mode the sensor measures the RF power
when a trigger signal is applied, and in the Delay Trig’d mode the sensor will sample the RF power
at some specific time later than the trigger.
See the Operation & Maintenance Manual included with the Triggerable Pulse (Peak Power) Sen-
sors for complete operating information.
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...