8003 Precision Scalar Analyzer
2-38
Manual 20791, Rev. C, June 2001
[
CAL SENSOR]
Menu F4:
Sensors and bridges must be calibrated before they can be used with the analyzer. The
calibration linearizes the output voltage versus power input characteristics of the diode sensors.
Once a sensor is calibrated, the analyzer will remember the calibration even when turned off as
long as the sensor is used on the same channel on which it was calibrated.
Initiates the calibration of the 8003’s sensors. When [
CAL SENSOR
] is pressed, the softkey menu
will change to F4-4 for selection of either all sensors or the single sensor of interest.
Make sure that the sensor or bridge to be calibrated is attached to the calibrator output on the front
panel of the 8003 before pressing any of the softkeys shown in menu F4-4.
[
CAL ALL]
Menu F4-4:
Initiates the routine to calibrate all of the 8003 sensors and bridges. The instru-
ment will test its sensors to try and determine which one is currently connected to the calibra-
tor, and then attempt to calibrate it. When it cannot calibrate a sensor, it will display a prompt
to connect a sensor to the calibrator.
[SENSOR (A, B, or C)
CAL]
Menu F4-4:
Initiates the calibration of the corresponding sensor.
While the 8003 is calibrating a sensor, a special calibration display will be active (see
Figure 2-21). This display shows the progress and success of the sensor calibration in progress.
If the instrument is unable to successfully complete the calibration (for example, the sensor/
bridge might be damaged), the 8003 will exit the calibration routine and generate an audible
warning signal.
During a sensor calibration, the softkey menu will offer the option to either Pause Cal or Abort
Cal (as described below).
When the calibration completes, the 8003 will return to the Cal key menu (F4).
[
PAUSE CAL
] causes the 8003 to pause while performing its sensor calibration function. This
can be done to allow an inspection of the calibration display. The first softkey menu prompt
will then change from Pause Cal to Cont Cal.
[
CONT CAL
] commands the 8003 to continue with its calibration.
[
ABORT CAL
] causes the 8003 to stop any calibration procedure which is currently being per-
formed, and return the routine to the Cal key menu (F4-1). Any previous calibration is
retained and used. The new, partial calibration data is eliminated.
Содержание 8003
Страница 3: ......
Страница 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Страница 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Страница 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Страница 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Страница 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Страница 38: ...8003 Precision Scalar Analyzer 2 2 Manual 20791 Rev C June 2001 Figure 2 1 8003 Front Rear Panel Components ...
Страница 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Страница 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Страница 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Страница 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...