SCPI Command Set
2500A Series Operation Manual, 34172 Revision C, March 2008
3.2.2.5
Status Subsystem
Table 3-8: Status Subsystem SCPI Commands
3.2.2.6
System Subsystem
[SOURce]:LIST:SYNC <sync>
Sets sync out option to <sync>. The sync out option determines how a
pulse is emitted from the SYNC OUT connector during List
operation. The choices are as follows:
0 No pulses are emitted from the SYNC OUT connector during List
operation.
1 A pulse is emitted from the SYNC OUT connector when the first
list point is executed.
2 A pulse is emitted from the SYNC OUT connector when the last
list point is executed.
3 A pulse is emitted from the SYNC OUT connector when each
point in the list is executed.
Command Syntax
Description
STATus:QUEStionable:CONDition?
Returns the value of the Questionable Status Condition Register.
The value returned is a decimal value representing the current state
of the register.
STATus:QUEStionable:ENABle <ques>
Sets the Questionable Status Enable Register.
Range of <ques> is 0 - 65535
Table 3-9: System Subsystem SCPI Commands
Command Syntax
Description
SYSTem:COMMunicate:GPIB[:SELF]:AD
DRess <address>|MAXimum|MINimum
Sets the instrument's GPIB address. The choices are as follows:
<address> Any integer between 1 and 30.
MAXimum Sets the GPIB address to 30.
MINimum Sets the GPIB address to 1.
SYSTem:COMMunicate:GPIB[:SELF]:AD
DRess?
Queries the instrument's GPIB address.
SYSTem:COMMunicate:SERial:BAUD
<rate>
Sets the RS-232 interface baud rate. The supported values for
<rate> are 9600, 19200, 38400, and 115200.
SYSTem:COMMunicate:SERial:BAUD?
Queries the current RS-232 interface baud rate.
SYSTem:COMMunicate:SERial:BITS
<bits>
Sets the number of RS-232 interface data bits. The supported values
for <bits> are 7 and 8.
SYSTem:COMMunicate:SERial:BITS?
Queries the number of RS-232 interface data bits.
SYSTem:COMMunicate:SERial:PARity
[:TYPE] EVEN | ODD | NONE
Sets the RS-232 interface parity type. The choices are as follows:
EVEN Selects even parity.
ODD Selects odd parity.
NONE Parity is not used.
SYSTem:COMMunicate:SERial:PARity?
Queries the RS-232 interface parity setting.
SYSTem:COMMunicate:SERial:SBITS
<sbits>
Sets the number of RS-232 interface stop bits. The supported values
for <sbits> are 1 and 2.
Table 3-7: Source Subsystem - List Mode SCPI Commands
Command Syntax
Description
Содержание 2508A
Страница 1: ...2500A Series 2500A Series Microwave Synthesizer Operations Manual 34172 Revision C March 2008 ...
Страница 4: ......
Страница 22: ...Chapter 1 2500A Introduction 22 2500A Series Operation Manual 34172 Revision C March 2008 ...
Страница 88: ...2500A Chapter 3 Remote Operation 88 2500A Series Operation Manual 34172 Revision C March 2008 ...
Страница 122: ...Chapter 4 Specifications Performance Verification 122 2500A Series Operation Manual 34172 Revision C March 2008 ...
Страница 132: ...2500A Series Microwave Synthesizer Error Messages 132 2500A Series Operation Manual 34172 Revision C March 2008 ...