INSTRUCTION MANUAL CENT series balances 60.50.670 3rd issue july 2005
Page 38 di 40
Status information of the F1 character
D
Transmitted data is valid
O
Balance in over range
U
Balance in under range
T
Tare
C
Calibration process in progress
Z
Zero acquisition
I
Init test
Status information of the F2 character
S
Measure is stable
I
Measure is not stable
E
Error
A
Automatic calibration (when F1 = C)
L
Load weight(manual calibration) (when F1 = C)
U
UnLoad weight (manual calibration) (when F1 = C)
D
Calibration done (when F1 = C)
B
Calibration busy(already started)
O
Calibration OFF
P
Percentage
11.2.2 EURO protocol
The answer to any type of command is formatted as follow:
D0 D1 D2 D3 D4 D5 D6 D7 D8 CR LF
1 2 3 4 5 6 7 8 9 10 11
The measure field (1..9) is the display reading.
It is right justified with decimal point and the measure sign.
Field N°10 and 11 are:
CR
carriage returns ASCII code.
LF
line feed ASCII code.
11.2.3 PRINTER protocol
Is activated by the “PRINT” pushbutton. The output string uses 10 characters for the measure including sign
and decimal point, the value is right justified and filled with spaces from the left. The format of the output is:
D0 D1 D2 D3 D4 D5 D6 D7 D8 D9 B U U U CR LF
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
The measure field (1..10) is the display reading.
It is right justified and has the decimal point and the measure sign.
Unit field shows the current unit weight
B is the blank character.