7 | Operation and maintenance
GMH 3331
24 / 32
B-H77.0.01.DB2-1.0
NOTE
Adjustment can be interrupted at any time with the set button. In this case, the existing
adjustment remains in effect.
If you press and hold the
cal
button for 10 seconds, the device will be reset to the
factor adjustment.
1. Press and hold the
cal
button for 2 seconds and then press the cal button again so
that
PT
. 1 appears in the secondary display.
2. The value of the buffer is adjusted with the with the
max
button and the
min
button.
Observe the permissible temperature compensation for the adjustment in the
chapter
General information for adjustment [
3. A valid, stable value has not been recognised as long as the display is blinking.
Observe the stability criteria in the chapter
General information for adjustment
4. Once the display no longer blinks, a valid and stable value has been found and
can be adopted with the
store
button.
5. If the error message
ERR9
briefly appears, observe the permissible sensor devi-
General information for adjustment [
.
6. If no error message appears, the value has been adopted successfully and the
second adjustment point can be entered.
7.
PT
. 2 appears in the secondary display. Now insert the TFS0100E sensor into the
adjustment device for the second humidity value. Observe the permissible adjust-
ment points in the chapter
General information for adjustment [
.
8. The value is adjusted with the with the
max
button and the
min
button. Observe the
permissible temperature compensation for the adjustment in the chapter
9. A valid, stable value has not been recognised as long as the display is blinking.
Observe the stability criteria in the chapter
General information for adjustment
10. Once the display no longer blinks and no longer switches, a valid and stable value
has been found and can be adopted with the
store
button.
11. If the error message
ERR9
briefly appears, observe the permissible sensor devi-
General information for adjustment [
.
12. If no error message appears, the value has been adopted successfully. The
product restarts and switches to normal operating mode.
Outcome of an action
The product is now adjusted and ready for measurement.
7.4 Certificates
The certificates are categorised as ISO calibration certificates and DAkkS calibration
certificates. The purpose of the calibration is to verify the precision of the measuring
device by comparing it with a traceable reference.
NOTE
The ISO standard 9001 is applied for the calibration certificates. These certificates
area affordable alternative to the DAkkS calibration certificates and provide informa-
tion of the traceable reference, a list of individual values and documentation.