10 System messages 

Er. 1  = 

measuring range has been exceeded, measured value is to high 

Er. 2  = 

measuring range is undershot, measured value is to low 

-- 

Sensor error: Value could not be calculated

 

Er. 7  = 

System fault - the device has detected a system fault (defective or far outside 

allowable ambient temperature range) 
If  

"BAT"

 is displayed at the left side of display, the battery is weak, measuring can be continued 

for a short period. 
If  

“bAt”

 is displayed in the main display the battery is used up and needs to be replaced. 

Measuring is no more possible. 

11 The serial interface 

11.1  Interface 

By means of the serial interface and a suitable electrically isolated interface adapter (USB 3100, USB 3100 
N, GRS 3100 or GRS 3105) the device can be connected to a computer for data transfer.  
With the GRS 3105 up to 5 devices of the GMH3xxx- series can be connected to one interface (see also 
manual of GRS 3105). As a precondition the base addresses of all devices must not be identical, make 
sure to configure the base addresses accordingly (refer menu point “Adr.” in chapter 7). 
To avoid transmission errors, there are several security checks implemented e.g. CRC. 
The following standard software packages are available: 

 

GMHKonfig: 

 

Software for a comfortable editing of the device (e.g. Material selection…) 

 

EBS 20M / 60M:

 

20-/60-channel software to display the measuring values 

In case you want to develop your own software we offer a GMH3000-development package including: 
• 

a universally applicable Windows functions library ('GMH3000.DLL') with documentation that can be 

used by the most programming languages. Suitable for Windows XP™, Windows Vista™, Windows 7™, 
Windows 8/8.1™, Windows 10™ 
• 

Programming examples Visua

l Basic 4.0™, Delphi 1.0™ 

 

The device has 2 channels: 

Channel 1: CO-concentration 

[ppm], [mg/m³] or [%COHb] 

 

Channel 2: temperature (nearly) 

[°C] 

 

Note: 

 The measuring-/ alarm- and display range values read back from the interface 

are always in the selected measurement unit! 

 
11.2  Supported interface functions 
 

Code  name/function 

code 

name/function 

code  name/function 

Read measurement value 

178 

Read measuring range unit 

204 

Read display decimal point 

Read system state 

179 

Read meas. range decimal 
point 

208 

Read number of channels 

Read max. measured value 

180 

Read kind of measuring or 
sensor 

222 

Read power-off time 

12 

Read ID number 

199 

Read kind of measuring or 
display 

223 

Set power-off time 

23 

Read max alarm limit 

200 

Read min. display range 

240 

Reset 

176 

Read min. measuring range 

201 

Read max. display range 

254 

Read Program version 

177 

Read max. measuring range  202 

Read display unit 

 

 

 

Note for operation with GAM3000:   

Keep in mind that the instrument just support max-, not min-alarm !  
Therefore the control-output of the GAM3000 could not be used, just the alarm output! 

 

 

Содержание GREISINGER GCO 100

Страница 1: ... 49 9402 9383 0 info greisinger de www greisinger de Operating Manual Carbon Monoxide CO Measuring Instrument with alarm function as of version 1 4 GCO 100 Please carefully read these instructions before use WEEE Reg Nr DE 93889386 Please consider the safety instructions Please keep for future reference ...

Страница 2: ...HE SERIAL INTERFACE 11 11 1 INTERFACE 11 11 2 SUPPORTED INTERFACE FUNCTIONS 11 12 SPECIFICATION 12 1 General Advice Read through this document attentively and make yourself familiar to the operation of the device before you use it Keep this document in a ready to hand way in order to be able to look up in the case of doubt The manufacturer will assume no liability or warranty in case of usage for ...

Страница 3: ...gnificant damage to property at non observance Attention This symbol warns of possible dangers or dangerous situations which can provoke damage to the device or environment at non observance Note This symbol point out processes which can indirectly influence operation or provoke unforeseen reactions at non observance 2 3 Safety guidelines This device has been designed and tested in accordance with...

Страница 4: ...swallowing give large volumes of water DO NOT induce vomiting Obtain medical treatment 3 Product Specification 3 1 Scope of supply Device GCO 100 incl 9V battery block Operation manual Calibration protocol 3 2 Operating and maintenance advice Battery operation If bAt is shown in the lower display the battery has been used up and needs to be replaced However the device will operate correctly for a ...

Страница 5: ...HLD Measured value is frozen key 3 4 2 Pushbuttons Left key on off button for power off press longer Middle key Max display of the maximum measured value MAX arrow in display Right key Hold hold frozen of the current measure value HLD in display 4 3 Connections 1 Interface Connect to optically isolated interface adapter accessory GRS 3100 USB3100 2 Sensor opening The mains socket is located at the...

Страница 6: ... the manufacturer have to be free of any residual of measuring media and other hazardous substances Measuring residuals at housing or sensor may be a risk for persons or environment Use an adequate transport package for reshipment especially for fully functional devices Please make sure that the device is protected in the package by enough packing materials 5 2 Disposal instructions Batteries must...

Страница 7: ...pm Strong smoker 20ppm Like with alcohol concentration in blood also the CO is decomposed in the blood around every 5h the CO content is cut by half Via the CO concentration of exhaled air the saturation of the CO in blood haemoglobin can be derived The carboxy haemoglobin in can be displayed directly by the instrument COHb p r t Configuration of the instrument This value is an estimation of the c...

Страница 8: ...nit mg m3 Selection of the display unit COHb AL Alarm function OFF No alarm ON Alerting over display and buzzer factory setting NO SO Alerting only via display AL K Alarm boundary not available if alarm function is deactivated AL oFF 0 1000 Alarm boundary in the selected display unit factory setting 10ppm RAIL Warning boundary for maximum working place concentration 0 100 Value for warning boundar...

Страница 9: ...ment Switch on and press Hold key during the segment test until the display shows the first parameter OFFS Press up or down key the currently selected offset adjustment appears Choose the desired value by pressing up or down key max selectable values 20ppm Enter by pressing On Off key OFFS appears in the display again Selsect the next parameter by pressing On Off key SCAL appears in the display Pr...

Страница 10: ... test gas cap GZ 10 to the instrument Connect the test gas cap GZ 10 to the valve unit via the hose The Gas bottle has to stand upright for the flow display of the gas valve unit is able to work properly Scale Adjustment Keep bottle upright open valve slowly until the ball in the flow indicator shows 0 5 l min lowest marking of indicator Wait until the measurement display at has stabilised ca 1min...

Страница 11: ...ng of the device e g Material selection EBS 20M 60M 20 60 channel software to display the measuring values In case you want to develop your own software we offer a GMH3000 development package including a universally applicable Windows functions library GMH3000 DLL with documentation that can be used by the most programming languages Suitable for Windows XP Windows Vista Windows 7 Windows 8 8 1 Win...

Страница 12: ...time 0 90 RH Storage temperature 10 50 C Housing impact resistant ABS membrane keyboard transparent panel Front side IP65 integrated pop up clip for table top or suspended use Dimensions 142 x 71 x 26 mm L x W x D Wight approx 155 g Power supply 9V battery in scope of supply as well as additional d c connector diameter of internal pin 1 9 mm for external 10 5 12V direct voltage supply suitable pow...

Отзывы: