5 | Operation and maintenance
G 1910 series
6.
You can change the value of the gradient compensation with the
up key
and
down
key
. Otherwise, confirm the preadjusted value with the
function key
. The 1-point
calibration is finished.
NOTE
Current test gases normally have accuracies of ± 2 %. This tolerance must be taken
into consideration with the measurement uncertainty. The specifications on the analy-
sis certificate must always be observed.
NOTE
For information about the available calibration settings, refer to Configuring parameters
of the configuration menu.
After the calibration is finished
[AL DONE
is displayed.
Then, the current measurement is shown in the display again.
If the calibration is not completed successfully an error message is displayed.
[AL ERR.
. Confirm the error mes-
sage pressing the
Function key
. The product restarts. The values of the last correctly
performed calibration are restored.