G 1111
Measurement Basics | 6
B-H88.0.11.DB2-1.0
17 / 28
6.3.2
A
V
R
0
:
02
/
A
V
R
0
:
05
/
A
V
R
0
:
10
Fast measurement with mean value over 2 s / 5 s / 10 s
Mean value mode for measurement of heavily fluctuating pressures.
_
Prerequisite
In the
Configuration mode
, a special function
A
V
R
0
:
02
,
A
V
R
0
:
05
or
A
V
R
0
:
10
has been
selected.
By pressing and holding the
Function key
for 4 s. the measurement with mean value
can be activated.
The different mean value times of 2, 5 or 10 seconds can be selected depending on
the requirement.
_
The first parameter is shown in the auxiliary display.
_
_
Special case
If the Tare function is activated when called up, this special function
A
V
R
can be reset
by pressing and holding the
Function key
for 4 s. In order to reactivate the Tare, the
special function must be switched in the configuration menu.
6.4 Pressure connections
_
NOTE
Suitable hoses must be used for vacuum measurements, silicon is suitable as a hose
material due to the low rigidity.
6.5 UT operation
Universal hose connection for 6 x 1 mm (4 mm inside) and 8 x 1 mm (6 mm inside)
hoses.
Fig. 5: UT connection
The universal hose connection is suitable for plastic and silicon hoses with an outer
diameter of 6 mm, for which purpose the hoses are simply connected to the upper
part. Rubber/silicon hoses with a larger diameter (e.g. 8 mm) can also be connected.
They are fit on the lower part for this purpose.
6.6 QC6 operation
Quick-Connect for Ø 6 mm hoses.
Fig. 6: QC6 Quick-Connect