5 | Measurement Basics

 

G 1110 

 

12 / 20 

 

B-H88.0.01.DK2-1.2 

Measurement Basics 

5.1 

General information about absolute pressure measurement 

The device measures absolute pressure. However, this should not be confused with 
the "sea level air pressure" indicated by weather stations. The altitude-based air 
pressure decrease is calculated for these pressure specifications.  
The device is capable of correcting this air pressure altitude correction. 

To do this, set the parameter value 

SEA.L 

in the

 Configuration menu

 to 

YES

 and enter 

the current altitude above sea level in the parameter value 

ALT,.

 

The nautical norm correction must be deactivated for vacuum measurements. 

5.2 

Altimeter 

If metres or feet are selected as the display unit in the 

Configuration menu

, the device 

is in altimeter mode. 
The unit calculates the current altitude from the current air pressure: At higher alti-
tudes, the air pressure is lower. It is important to note that not only changes in altitude 
but also the weather influence the measurement. To correct this weather influence, the 
displayed altitude can be corrected via the keys. 

Altimeter display: 

In the Altimeter operating mode, you will be prompted to enter the 
current altitude after pressing the 

Function key

 for 4 s. 

Note:   The first time you are prompted after switching on the device, the value is set to 

the alti value in the 

Configuration menu

. (default value for starting point) 

The current value is retained for all further calls. 

Note on min/max display:   in the secondary display, the sum of the altitude metres 

covered is also shown 
for min: dESC - descent, for max: ASC - ascent 

5.3 

Special functions 

5.3.1 

NVLL 

Tare function 

If a pressure unit is selected as the display unit in the 

Configuration menu

, the device 

is in barometer mode.  
After it is switched on, the device starts up in standard measuring mode, the special 
function “tare” is started by pressing and holding the 

Function key

 for 4 s.. 

Barometer display:  The display can be zeroed by pressing the 

Function key

 for 4 s.  

If the tare function is activated, 

NVLL

 blinks in the lower display. 

The tare function can be reset by pressing the 

Function key

 again 

for 4 s.  

Содержание GREISINGER G 1110

Страница 1: ...Quick reference guide G 1110 barometer altimeter EN ...

Страница 2: ... glance 6 3 1 Display elements 6 3 2 Operating elements 7 4 Operation 8 4 1 Opening the configuration menu 8 4 2 Open the adjustment menu 11 5 Measurement Basics 12 5 1 General information about absolute pressure measurement 12 5 2 Altimeter 12 5 3 Special functions 12 5 3 1 NVLL Tare function 12 5 4 Use of the trend indicator 13 5 4 1 Meteorology weather forecasting 13 5 4 2 Hiking cycling flying...

Страница 3: ...and consequential damag es are voided with misuse disregarding this document disregarding safety notices assignment of inadequately qualified technical personnel and arbitrary modifications of the product This document is entrusted to the recipient for personal use only Any impermissible transfer duplication translation into other languages or excerpts from this operating manual are prohibited The...

Страница 4: ...anteed if applicable safety precautions and the device specific safety instructions for this docu ment are observed If these notices are disregarded personal injury or death as well as property damage can occur DANGER Incorrect area of application In order to prevent erratic behaviour of the product personal injury and property damage the product must be used exclusively as described in the chapte...

Страница 5: ...es for which it was designed It must be handled with care and used according to the technical data do not throw strike etc Suitable measures must be used to protect the pressure equalisation opening and be protected from dirt 2 5 Qualified personnel For commissioning operation and maintenance the relevant personnel must have adequate knowledge of the measuring process and the significance of the m...

Страница 6: ... _ Battery indicator Evaluation of the battery status _ Unit display Display of the units or Min Max Hold information text _ Main display Measurement of the current pressure or value for min max hold _ Auxiliary display Measurement of the current pressure in Min Max Hold mode _ Bar display Trend display in 7 steps falling strong medium slight stable rising slight medium strong _ ...

Страница 7: ...in max value Change value of the selected parameter Long press Reset the min max value of the current measure ment Both simultaneously Rotate display overhead display _ Function button Press briefly Freeze measurement Hold Call up next parameter Long press 2s Start menu configuration ONF appears in the display Long press 4s Activation of the Tare function NVLL altitude input Operating status devic...

Страница 8: ...nit HPa Barometer in hPa mBaR Barometer in mbar PSI Barometer in psi mmKG Barometer in mmHg m Altimeter Elevation display in m FT Altimeter Elevation display in feet __RATE Measuring rate SLO Slow measurement speed FAST Fast measurement speed not recommended for altimeter display __SEA L SeaLevel correction not available at altimeter elevation display NO Inactive display measured air pressure dire...

Страница 9: ... at sea level I_H 3_H Time base for bar display 1 hour 1 hPa bar level or 3 hours 3 hPa bar level for usual meteorologi cal assessments The value will be renew every minute _LCD 2 Additional information OFF No additional information in auxiliary display Temperature in C F Temperature in F With activated TEND display additionally TEND Trend value TE Temperature in C and trend value F TE Temperature...

Страница 10: ...ve been pressed L TE Backlight OFF Backlight deactivated 0 15 0 30 1 00 2 00 4 00 Automatic shut off of the backlight after a selected time in minutes seconds during which no buttons have been pressed ON No automatic shut off of the backlight IN T Factory settings NO Use current configuration YES Reset device to factory settings After confirming with the function button the display shows IN T DONE...

Страница 11: ... the down button and press the On Off button to switch on the device and open the Adjustment menu 3 The display shows the first parameter Release the down button Parameter Values Meaning _PR OF Zero point correction 0 0 No zero point correction 5 0 5 0 Zero point correction in selected display unit e g 5 0 hPa PR SL Gradient correction 0 00 No gradient correction of the temperature 5 00 5 00 Gradi...

Страница 12: ...lso the weather influence the measurement To correct this weather influence the displayed altitude can be corrected via the keys Altimeter display In the Altimeter operating mode you will be prompted to enter the current altitude after pressing the Function key for 4 s Note The first time you are prompted after switching on the device the value is set to the alti value in the Configuration menu de...

Страница 13: ...efore not shown in the display before this time has elapsed 5 4 1 Meteorology weather forecasting Observation of variable weather conditions by assessing the rate of change of air pressure To do this set the device to UN T HPa international standard RATE SLO power saving during continuous operation SEA L YES ALT altitude of the location above sea level TEND 1_H or 3_H 1 hour 1 hPa bar level or 3 h...

Страница 14: ...d to the selected setting In the 60_S setting the resolution is reduced to 1 m or 5 ft To do this set the device to Hiking cycling flying motor sports UN T m FT oder m Altimeter RATE SLO FAST power saving during continuous operation fast measured value response 1 ALT default value for input start TEND 1_H 1_S oder 60_S time base 1 second for m s or 60 seconds for ft min LCD 2 TEND TEND or for more...

Страница 15: ...ully depleted 6 1 2 Changing battery _ DANGER Danger of explosion Using damaged or unsuitable batteries can generate heat which can cause the batter ies to crack and possibly explode Only use high quality and suitable alkaline batteries CAUTION Damage If the batteries have different charge levels leaks and thus damage to the device can occur Only use high quality and suitable alkaline batteries Do...

Страница 16: ...nded depth 4 Fit the cover D on evenly The O ring must remain at the intended depth 5 Tighten the Phillips screws A 7 Disposal Separation by material and recycling of device components and packaging must take place at the time of disposal The valid regional statutory regulations and directives applicable at the time must be observed _ NOTE The device must not be disposed of with household waste Re...

Страница 17: ...r repair BAT Battery depleted Battery depleted Replace battery ERR 1 Measuring range exceeded Measurement too high Product is defective Stay within allowable measurement range Send in for repair ERR 2 Measuring range is undercut Measurement too low Product is defective Stay within allowable measurement range Send in for repair ERR 3 Display range has been exceeded Incorrect display unit Value not ...

Страница 18: ... Measuring medium air Pressure connection No connection integrated sensor Pressure equalisation via diaphragm protected opening Measuring cycle FAST approx 2 5 measurements per second SLO approx 4 seconds Display 3 line segment LCD additional symbols illuminated adjustable white permanent illumination Standard function Min Max Hold Trend display time base selectable 1s 60s 1h 3h corresponds to Alt...

Страница 19: ...lighting and with measuring rate Slo Battery indicator 4 stage battery status indicator Replacement indicator for depleted batteries BAT Auto power OFF function The device switches off automatically if this is activat ed Directives and standards The devices conform to the following Directives of the Council for the harmonisation of legal regulations of the Member States 2014 30 EU EMC Directive 20...

Страница 20: ... 10 Service 10 1 Manufacturer If you have any questions please do not hesitate to contact us GHM Messtechnik GmbH GHM GROUP Greisinger Hans Sachs Str 26 93128 Regenstauf GERMANY Email info greisinger de www greisinger de WEEE reg no DE 93889386 ...

Отзывы: