E40.0.01.6C-08 

Operating Manual EB-Logg 80CL 

page 8 of 16 

5.2  Operating possibilities at the device - logger operation 

Function 

Procedure 

Display and adjust 
cycle time 

Push button 

 and 

 at the same time. 

The main display shows “CYCL”. The secondary display 
shows the currently adjusted cycle time [in sec.]. 

If the logger is currently not recording, the desired cycle time 
[4 sec to 5h (18000 sec)] can be adjusted via the buttons 

 

 and 

 . 

Press button  

 again to confirm your adjustments and to 

leave the menu. 

Note:  

If no key is pressed for more than 2 min, the menu is 
quit without saving.  

Start recording 

Push button 

 for about 5 seconds. 

The main display shows “run”.  
The secondary 

display shows “no”

To start recording select “YES” with the arrow keys 
(button 

 or 

 ) and verify with 

 . 

Note: 

If no key is pressed for more than 2 min, the menu is 
quit without saving. 

Stop recording 

Push button 

 for about 5 seconds. 

The main 

display shows “HoLd”.  

The secondary display shows “no”
To stop recording select “YES” with the arrow key 
 (button 

 or 

 ) and verify with 

 . 

Note: 

If no key is pressed for more than 2 min, the menu is 
quit without saving. 

Notice: 

If you have activated one of the following options  
 

• “logger stop via buttons disabled”  

 

• “do not stop active logger”  

at the configuration of the logger the menu to 
stop the recording cannot be called. 

 

SET

1

2

2

3

SET

1

SET

1

2

3

SET

1

SET

1

2

3

SET

1

Содержание GREISINGER EB-Logg 80CL

Страница 1: ...ik GmbH Hans Sachs Str 26 93128 Regenstauf GERMANY Tel 49 9402 9383 0 info greisinger de www greisinger de from from Version V2 7 Operating Manual EB Logg 80CL EB Logg 80CL E Data logger for humidity...

Страница 2: ...Operating possibilities via software EASYBus Configurator 10 6 Battery life time and recording period 10 7 Operating displays 11 7 1 Measured value display 11 7 2 Logger status messages 11 7 3 System...

Страница 3: ...ge This device must not be used at a patient for diagnostic or other medical purpose The liability and warranty of the manufacturer for damages and consequential damages are voided with misuse disrega...

Страница 4: ...only be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification 5 When connecting the device to other devices e g PC the interconnection has to...

Страница 5: ...tings 3 2 Notes on the logger s delivery status The delivered logger is in a sleep mode that means nothing is displayed and the power consumption is at its minimum When there is a communication with t...

Страница 6: ...logger is recording BAT signals weak battery see chapter 7 AL signals an alarm state at minimum one channel SL signals an active sea level correction for the air pressure 4 2 Function of the buttons b...

Страница 7: ...the display el ement MAX Note You can immediately switch to the next measuring channel pushing the button again Delete max value Push button for about 5 seconds CLr is displayed and the max values of...

Страница 8: ...tart recording Push button for about 5 seconds The main display shows run The secondary display shows no To start recording select YES with the arrow keys button or and verify with Note If no key is p...

Страница 9: ...rm your adjustments and to leave the menu Note If no key is pressed for more than 2 min the menu is quit without saving This setting allows the logger to be a substitute for outdated loggers in existi...

Страница 10: ...an be only changed if the logger is stopped logger without logger data 5 5 Operating possibilities via software EASYBus Configurator The software EASYBus Configurator V2 0 is needed to change the conf...

Страница 11: ...row AL signals an alarm state at minimum one channel The arrow SL signals an active sea level correction 7 2 Logger status messages At stand alone operation no EASYBus connected logger status messages...

Страница 12: ...repair 7 4 Alarm and error messages The following messages correspond to a particular measuring channel and are dis played accordingly ALARM LOW The measured value is below the selected min alarm lim...

Страница 13: ...RH Air pressure 1 0 mbar typ 2 5 mbar max Display two 4 digit LCD displays about 7 and 5 8 mm high Interface EASYBus Busload equals to 2 EASYBus elements Battery 1x fixed lithium cell Li SOCl2 3 6V 1...

Страница 14: ...lowing European di rectives 2011 65 EU RoHS 2014 30 EU EMV directives Applied harmonized standards EN 50581 2012 EN 61326 1 2013 emissions level class B emi immunity according to table 1 additional er...

Страница 15: ...delete data Problem A error message is displayed Error message Possible reason procedure remedy measuring range exceeded Measuring range of the chan nel has been exceeded Operate logger within its sp...

Страница 16: ...es a General connection diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be connected to terminal 11 of the GNR NPN to be connec...