3. 

Product description 

3.1  Required accessory 

The logger is configured, started and read-out via the 

EASYBus

-interface.  

Therefore the following accessories are required:  

 

Level converter

 

- RS232 <> 

EASYBus

 

(i.e. EBW 1, EBW 64, EBW 240) 

- USB <> 

EASYBus

 

(i.e. EBW 3) 

  Connecting cable: level converter to 

EB-Logg

 

 

GSOFT 40K

 (version >7.14): 

 

  Windows software to start the logger and read out the logger-data 

 

EASYBus-Configurator

 (version >2.0): 

  for the configuration of the extended settings 

3.2 

Notes on the logger’s delivery status 

The delivered logger is in a “sleep mode”, that means nothing is displayed and the power 
consumption is at its minimum. When there is a communication with the software for the 
first time, the EB-Logg 

leaves the sleep mode and ›Stop‹ is displayed. Then the device is 

ready for operation. 

Note: 

That sleep mode (nothing is displayed) cannot be restored again. 
“Stop“ or  “Halt“ are similar to the sleep mode. Both of that operation 
modes ensure minimum consumption. 

Note: 

Logger start over keypad: 
before logger start over keypad, logger time must be adjust with 
GSOFT 40K. 

3.3  Connection 

 

 

EASYBus 

assignment of jack 

(view on pins) 

 

EASYBus 

 
 
 

 

not used 

 

 
 
 
 
 
Sensor 

or cable with sensor 
     (EB-Logg 80 KH) 
or sensor connection 
     

(with option “-AFK”) 

 

 

 

 

Содержание GREISINGER EB log 80 K SMB2

Страница 1: ...Y Tel 49 9402 9383 0 info greisinger de www greisinger de Installation and operating manual Data logger for temperature from version 1 0 EB Logg 80K Please read these instructions carefully before use...

Страница 2: ...ASYBus Configurator 8 7 Battery life time and recording period 8 8 Operating displays 9 8 1 Measured value display 9 8 2 Logger status messages 9 8 3 System messages 10 8 4 Alarm and error messages 10...

Страница 3: ...ufacturer for damages and consequential damages are voided with misuse disregard of this operating manual assignment of inadequately qualified technical personnel and arbitrary modifications of the pr...

Страница 4: ...nly be guaranteed if the device is not subjected to any other climatic conditions than those stated under Specification 5 When connecting the device to other devices e g PC the interconnection has to...

Страница 5: ...he logger s delivery status The delivered logger is in a sleep mode that means nothing is displayed and the power consumption is at its minimum When there is a communication with the software for the...

Страница 6: ...ng attached on the device EB Logg 80KH sensor tube made of stainless steel 5 mm ap prox 50 mm long approx 1 m silicone cable with anti buckling glanding to housing EB Logg 80KH SMB1 sensor tube made o...

Страница 7: ...n again Delete min value Push button for about 5 seconds CLr is displayed and the min values of all measuring channels are deleted 6 2 Operating possibilities at the device logger operation Function P...

Страница 8: ...s and alarm delay 1 activate the function logger stop via buttons disabled 1 Please note The alarm settings and the sea level correction can be only changed if the logger is stopped logger without log...

Страница 9: ...atus messages are dis played STOP The logger recording has been stopped The logger memory is empty Note At this state the power consumption is at its minimum HOLD The logger recording has been halted...

Страница 10: ...end the logger to the manufacturer for repair 8 4 Alarm and error messages The following messages correspond to a particular measuring channel and are dis played accordingly ALARM LOW The measured val...

Страница 11: ...ttery life time depending on the chosen measuring cycle about 5 years at 15 minutes and nominal temperature Measured value memory 250 000 measuring values of each variable Measuring cycle 4 seconds to...

Страница 12: ...n AFK The sensor socket is ESD endangered when the sensor is disconnected Disconnecting the sensor in EASYBus mode is not permitted Maximum sensor cable length 3m 10 Decommissioning reshipment and dis...

Страница 13: ...The logger has still saved data Remedy Stop logger as recommended by the software delete data Problem A error message is displayed Error message Possible reason procedure remedy measuring range exceed...

Страница 14: ...ck view on pins EASYBus not used Connection examples a General connection diagram for a relay with self induction recuperation diode b Connection information for GR 10 GNR 10 or GNR 232 A GND to be co...

Отзывы: