Direct connection of 4 wire Pt100 sensors to a DIN45326 connector
4 wire Pt100
sensors
can
be soldered di-
rectly to the pins of the flying female
connector DIN45326 without making use
of the TP47 board. The 4 wires of the
Pt100 sensors have to be soldered as in-
dicated in the figure on the left. In order
to use this type of probe it is necessary to
set up the menu item “Probe Type” as
described at page 8. The P100 probe is
recognized upon turning on the instru-
ment: connect the probe when the in-
strument is switched off and then turn it
on. This connector does not guarantee the
IP67 protection degree.