HD404…
- 6 -
V1.2
TAB. 2.1: pressure measuring ranges, accuracy and long term stability
MODEL
MEASURING RANGE
ACCURACY
(1)
(@ 0…+50 °C)
LONG-TERM
STABILITY
(2)
LOW
HIGH
AZ
NO AZ
AZ
NO AZ
MODELS WITH ANALOG OUTPUT (HD404T…)
Pa (HD404TxP…)
HD404T1PGAZ…
0...50
0...100
±(0.8% measure
+ 0.5)
-
≤
±0.2
-
HD404T2PGAZ…
0...100
0...250
HD404T3PG…
0...250
0...500
±1% f.s. nom.
≤
±8
HD404T4PG…
0...500
0...1000
HD404T1PDAZ…
±50
±100
-
-
HD404T2PDAZ…
±100
±250
HD404T3PD…
±250
±500
±1% f.s. nom.
≤
±8
HD404T4PD…
±500
±1000
mmH
2
O (HD404TxM…)
HD404T1MGAZ…
0...5
0...10
±(0.8% measure
+ 0.05)
-
≤
±0.02
-
HD404T2MGAZ…
0...10
0...25
HD404T3MG…
0...25
0...50
±1% f.s. nom.
≤
±0.8
HD404T4MG…
0...50
0...100
HD404T1MDAZ…
±5
±10
-
-
HD404T2MDAZ…
±10
±25
HD404T3MD…
±25
±50
±1% f.s. nom.
≤
±0.8
HD404T4MD…
±50
±100
inchH
2
O (HD404TxI…)
HD404T1IGAZ…
0...0.2
0...0.4
±(0.8% measure
+ 0.002)
-
≤
±0.0008
-
HD404T2IGAZ…
0...0.4
0...1
HD404T3IG…
0...1
0...2
±1% f.s. nom.
≤
±0.04
HD404T4IG…
0...2
0...4
HD404T1IDAZ…
±0.2
±0.4
-
-
HD404T2IDAZ…
±0.4
±1
HD404T3ID…
±1
±2
±1% f.s. nom.
≤
±0.04
HD404T4ID…
±2
±4
MODELS WITH RS485 MODBUS-RTU OUTPUT (HD404ST…)
HD404ST2AZ…
±250 Pa
±(0.8% measure
+ 0.5) Pa
≤
±0.2 Pa
HD404ST4…
±1000 Pa
±1% f.s.
≤
±8 Pa
HD404ST5…
±100 mbar
±(0.8% measure
+ 0.005) mbar
±1% f.s.
≤
±0.002
mbar
≤
±0.08
mbar
(1)
f.s. nom. (nominal) = full scale of “HIGH” measuring range.
(2)
Long-term stability refers to 1 year.
Содержание Delta OHM HD404 Series
Страница 23: ......