10
8.
Software malfunction.
8.
Reset (chap. 8.7)
Individual motor does not work.
1.
Motor-plug is not connected
properly.
2.
MCP is defective.
3.
Control-box is defective.
4.
Motor-cable or plug is defective.
5.
Motor is defective.
1.
Connect plug properly.
2.
MCP must be replaced.
3.
Control-box must be replaced.
4.
Cord/plug must be replaced.
5.
Motor must be replaced.
15
TECHNICAL DATA
15
Characteristic
Value
Dimensions (length*width*height)
205*85*50 cm
Lying surface (length*width)
205*85 cm
Base (length*width)
approx. 132*60 cm
Height (no height adjustment)
approx. 49 cm
Inclination of headrest
0
0
to 40
0
Inclination of backrest
0
0
to 45
0
Inclination of legrest (thigh section)
0
0
to 23
0
Inclination of legrest
0
0
-30
0
Lying surface lifting capacity
(max. safe operational load)
200 kg (centrally-positioned load on
the centre of the lying surface; not
permitted on the head and foot side)
Backrest lifting capacity
(max. safe operational load)
75 kg
Total weight
approx. 90 kg
Adjustment drive incl. controls
Power of transformer (primary)
230V AC/50Hz/200 VA max.
Power of transformer (secondary)
24 V DC
Protection
3 ampere AT
Duty cycle
10 % max. 2 min. uninterrupted
operation/18 min. pause
Power cord
3 m, cord
MCP
Coiled cord
Device protection class
I
Overvoltage protection class
Type B
Protection against solid and liquid
materials
IP 54
Ambient conditions for storing and transportation
Temperature
-10
0
C to +50
0
C
Atmospheric pressure
700 to 1060 hPa
Relative humidity
20% to 90% at +30
0
C non-condensing