background image

2

8850-3 Conductivity/Resistivity Transmitter

14
13
12
11

Shield (Sensor Gnd)
Black (Iso Gnd)
White (Temp In)
Red (Signal In)

Signet 28XX-1

Standard and Certified Cells

Terminals

3.1  Sensor Input Connections

≤ 30 m (100 ft)

Signet  

Conductivity/Resistivity 

Transmitter 

ENTER

62.50 uS/cm

25.0 C

Sensr Gnd

(SHIELD)

Iso. Gnd

(BLACK)

Temp. IN

(WHITE)

Signal IN

(RED)

14

13

12

11

Output 2-

Output 2+

Output 1-

Output 1+

10

9

8

7

Loop 2-

Loop 2+

System Pwr

Loop -

System Pwr

Loop +

AUX

Power -

AUX

Power +

6

5

4

3

2

1

3. Electrical 

Connections

Caution:

  Failure to fully open terminal jaws before removing wire may permanently damage instrument.

Wiring Procedure
  1.  Remove 0.5 to 0.625 in. (13 to16 mm) of insulation from wire end.
  2.  Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
  3.  Insert exposed (non-insulated) wire end in terminal hole until it bottoms out. 
  4.  Release orange terminal lever to secure wire in place. Gently pull on each wire to ensure a good connection.

Wiring Removal Procedure
  1.  Press the orange terminal lever downward with a small screwdriver to open terminal jaws.
  2.  When fully open, remove wire from terminal.

Wiring Tips:

•  Do not route sensor cable in conduit containing AC power wiring. Electrical noise may interfere with sensor signal.
•  Routing sensor cable in grounded metal conduit will help prevent electrical noise and mechanical damage.
•  Seal cable entry points to prevent moisture damage.
•  Only one wire should be inserted into a terminal. Splice double wires outside the terminal.

2
1

Terminals 7-10: Open-collector Outputs

•  Two transistor outputs, programmable as:
 

• High or Low setpoint with adjustable hysteresis

 

• Proportional Pulse (to 400 pulses per minute maximum)

 

• May be disabled (Off) if not used

Terminals 3–6: Loop Power

12 to 24 VDC ±10% system power 

and current loop output.

Max. loop impedance:

  50 

 max. @ 12 V

325 

 max. @ 18 V

600 

 max. @ 24 V

Terminals 11–14: Sensor Input

11 is conductivity input
12 is  temperature input
13 is the isolated signal ground
14 is  the sensor earth ground

Terminals 1–2: Auxiliary power

Provides DC power to measurement 
circuit. Required for all 8850 systems

Internal open-collector 

output circuit 

Outputs 

Isolation 

15 Ω

 

 

Содержание Signet 8850-3

Страница 1: ...Maximum Altitude 2000 m 6562 ft Insulation Category II Pollution Degree 2 Standards and Approvals CE UL listed Manufactured under ISO 9001 for Quality ISO 14001 for Environmental Management and OHSAS...

Страница 2: ...rocedure 1 Press the orange terminal lever downward with a small screwdriver to open terminal jaws 2 When fully open remove wire from terminal Wiring Tips Do not route sensor cable in conduit containi...

Страница 3: ...Recorder Channel 3 4 to 20 mA in Channel 4 4 to 20 mA in Transmitter 1 Terminals Transmitter 2 Terminals PLC Terminals Channel 1 4 to 20 mA Power Supply 12 to 24 VDC Channel 2 4 to 20 mA Channel 2 4 t...

Страница 4: ...mple below The output will be 0 pulses min when value is less than 5 The output will be 50 pulses min when value is 7 5 The output will be 100 when value is greater than 10 The starting point endpoint...

Страница 5: ...OWN keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with UP or DOWN arrow keys Step 4 Press RIGHT ARROW key to select menu ite...

Страница 6: ...ct the source for the Open Collector output Conductivity or Temperature Select the mode of operation for the Open Collector output Options available are High Low or proportional Pulse The signal may b...

Страница 7: ...riences frequent or extreme fluctuations Adjust the minimum and maximum current output The display value represents the precise current output Adjustment limits 3 80 mA 4 00 mA 5 00 mA 19 00 mA 20 00...

Страница 8: ...t value If the instrument does not display the correct value Temperature 0 5 C Conductivity 2 of reading service is required A Accuracy Verification with Precision Resistors Electronic Calibration 1 S...

Страница 9: ...ture compensation factor by entering 0 00 2 Heat the sample solution close to the maximum process temperature Place sensor in the sample solution allowing several minutes for stabilization Access the...

Страница 10: ...k sensor for correct range Check Decimal setting in OPTIONS menu Check Calibrate menu settings for incompatible SOURCE and RANGE values Set a PPM value less than 3 Set cell constant to a value greater...

Страница 11: ...11 8850 3 Conductivity Resistivity Transmitter Notes...

Страница 12: ...4 Universal mounting kit 3 8050 395 159 000 186 Splashproof rear cover 3 8052 159 000 188 3 4 Integral mounting kit 3 8052 1 159 000 755 3 4 NPT mount junction box 3 0000 596 159 000 641 Heavy duty wa...

Отзывы: