GF Signet FlowtraMag 2580 Series Скачать руководство пользователя страница 11

11

Signet 2580 FlowtraMag Meter 

App Confi guration - Sensor Setup

Note:

If the GF Confi g Tool password has been lost or forgotten, connect 
blue wire to white wire while unit is powered (for 2 to 5 seconds.) 
Password will reset to factory original (last 6 digits of serial number.)

When the FlowtraMag Meter is in operation, when in close proximity to the FlowtraMag (less than 20 ft), open the 

GF Confi g Tool App

 to begin a search nearby devices and go thru the pairing process. Click on connect next to device 

you are pairing to.

1.  Pair the device by entering the device 

Code/Pin. 
The default Passkey is the last 6 digits of 
the product serial number.

2.  Click 

Pair/OK

3.  Make any adjustments to the FlowtraMag, 

if necessary, by tapping the Hamburger 
Menu (menu list) or Gear (edit settings).

Read

 

 Loads the data from the connected device (sensor or transmitter) and updates the software’s display. 

NOTE:

  This will overwrite any changes made in the GF Confi g Tool App since the last 

Write

.

Write

 

 Applies the data entered in the GF Confi g Tool App to the connected device. Once you have entered the desired setting 
changes in the software screens, press 

Write

 to load your new settings onto the connected device.

Save

 

 Stores the entire GF Confi g Tool App settings confi guration, as currently displayed in the application, to your mobile device. 
(You will be asked to provide a confi guration fi le name)

Load

 

 Loads a default confi guration fi le from the factory.
Opens a previously saved settings confi guration fi le. See 

Save

 function above.

NOTE:

   The fi le must be a GF Confi g Tool App settings confi guration fi le. 

The software will verify whether the user-selected fi le is the correct type.
There are confi guration fi les available for specifi c body sizes containing default values from the factory.
Carefully review the 

Device Tag

 and 

Passkey

 confi guration in the Information screen. 

Device Tag

 identifi es the sensor you are connecting to. Device Tag maximum length 20 characters. 

Device Passkey

 is needed for connecting to the sensor. Device Passkey is a 6 digit number.

Config Tool

iOS

Android

(Version 8 
or newer)

READ 

WRITE 

SAVE 

LOAD

Code/Pin

Hamburger
(Menu)

Gear
(Settings)

(example shows Menu)

Содержание Signet FlowtraMag 2580 Series

Страница 1: ...n the market It is designed for high accuracy flow measurement in short pipe runs making it an ideal solution for industrial applications where performance and ease of use are important The FlowtraMag is available in pipe sizes of 1 2 and 4 optimized for performance in short pipe runs often associated with final effluent lines well heads and water treatment skids Features include No moving parts No pr...

Страница 2: ...the plastic encloses other sensitive components as is the case with GF Signet FlowtraMag meter these may be affected or damaged by the media diffusing through the plastic body Unit is factory shipped configured to measure water 1 Depressurize and vent system prior to installation or removal 2 Confirm chemical compatibility before use 3 DO NOT exceed maximum temperature or pressure specs 4 ALWAYS wear ...

Страница 3: ...L Output Digital S3 L 4 5 to 5 5 VDC Serial ASCII TTL level 9600 bps Compatible with Signet 8900 9900 9950 and 0486 Profibus Concentrator Max Cable Length Application dependent The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by Georg Fischer is under license Other trademarks and trade names are those of their respective owners Sensor ...

Страница 4: ...in Union ends and union nuts shown 99 0 mm 3 90 in 276 86 mm 10 90 in 1 8 in 327 7 mm 12 90 in 99 0 mm 3 90 in 239 26 mm 9 42 in 1 16 in 249 4 mm 9 82in 141 2 mm 5 56 in 141 2 mm 5 56 in 99 0 mm 3 90 in Pipe 1 in ID Pipe 2 in ID Pipe 4 in ID 249 4 mm 9 82 in 31 8 mm 1 25 in 7 1 mm 0 28 in x 2 28 7 mm 1 13 in x 2 35 8 mm 1 41 in 199 64 mm 7 86 in 1 16 in 8 42 0 10 in Pipe cutout dimension 38 1 mm 1...

Страница 5: ...es A vacuum breaker may be required downstream of the FlowtraMag to ensure pipe doesn t drain and fill with air Sensor Pipe Installation 1 Choose a mounting location that satisfies the requirements 2 Install sensor with flow arrow pointing in the direction of flow Note Gland fittings should point upstream of flow OK Recommended OK Vertical flow is OK IF the pipe remains FULL at all times 2580 4 in Bolts...

Страница 6: ...the enclosure 3 2580 Black Red White Red Power cable Output cable Shield Black Brown Shield Blue used for GF Config App password reset only S3 L Frequency 4 to 20 mA Application Tip If your flow is in the reverse direction it is possible to set up reverse flow via the Signet 0252 Configuration Tool or GF Config Tool App 1 2 4 Units GPM GPM GPM Totalizer Units Gallons Gallons Gallons Temperature Units ...

Страница 7: ...RED Signet Flow Instruments CAUTION Turn off Power before Wiring DO NOT REMOVE Seal Warranty void if seal is broken or removed WARRANTY VOID IF REMOVED Black Red Shield EMC enhancement 24 VDC 1A min Power Supply Power cable WARRANTY VOID IF REMOVED 24 VDC 1A min Power Supply 2580 4 to 20 mA Black 4 to 20 mA Brown Output cable White Red Shield Blue not used Not used 2580 Power cable Black Red Shield...

Страница 8: ...ack Freq Input Red GND Shield 5VDC Black Freq Input 2 Red S L Red GND White Shield 5VDC Black S L Red GND White Shield 3 3 Analog Output 1 Analog Output 2 if applicable if applicable Frequency Input 1 Frequency Input 2 OR S3L Input 2 S3L Input 1 Frequency 1 S3 L 2 Ground White Data Red Output cable Black Brown Shield Blue not used Not used OR OR 2580 Ground White Data Red Output cable Black Brown ...

Страница 9: ...cy wiring and programing instructions Wiring with Digital S3 L Output PWR 9900 Power Input LOOP LOOP PWR 9900 S3 L Terminal DATA GND V SHLD 9900 Transmitter S3L Freq 24 VDC 1A min Power Supply 2580 Ground White Data Red Output cable Black Brown Shield Blue not used Not used 2580 Power cable Black Red Shield EMC enhancement 9950 Digital S3 L V FREQ DATA GND V FREQ DATA GND Loop1 Loop1 Loop2 Loop2 P...

Страница 10: ...Return to home screen and look for App icon click the blue GF Config Tool icon 3 Continue to Sensor Setup Section next page Bluetooth App Setup Steps Android version Download the GF Configuration Tool App by scanning the QR code or searching in Google Play directly 1 When prompted press Install 2 Return to home screen and look for App icon click the blue GF Config Tool icon 3 Continue to Sensor Setup...

Страница 11: ...ite Applies the data entered in the GF Config Tool App to the connected device Once you have entered the desired setting changes in the software screens press Write to load your new settings onto the connected device Save Stores the entire GF Config Tool App settings configuration as currently displayed in the application to your mobile device You will be asked to provide a configuration file name Load...

Страница 12: ...press WRITE to save your changes to the sensor Otherwise it saves to your phone only To switch between S3 L and Frequency and or 4 to 20 Active or Passive use the GF Config Tool App On the loop screen use the drop down to select S3 L or Freq and or Active or Passive 4 to 20 mA Press WRITE after making selection If the GF Config Tool password has been lost or forgotten connect blue wire to white wire...

Страница 13: ...raph span Low Flow Cutoff Factory setting is 0 02 m s 0 07 ft s equivalent flow rate and can be customized to any user preferences Averaging Time Factory setting is Low Can be customized Off Low Med High Sensitivity Factory setting is 0 5 m s 1 64 ft s equivalent flow rate and can be customized to user preferences Calibration Customization and Performance Settings Application Note The unit is factory ...

Страница 14: ... two tools that are designed to work around these deficiencies The Averaging and the Sensitivity features should be studied before making adjustments Averaging Time in Seconds Factory set Low Set the time the meter will use as the averaging period The ranges are Off Low 10 s Med 40 s and High 120 s Use higher averaging times to smooth the display and current output where the flow in the pipe is errat...

Страница 15: ...ng Sensitivity Low Flow Cut Off and Direction of the Flow 4 If the user has any correction or change you must Write the information to device then click Read device again 1 2 This is an outline For complete instructions please refer to the 0252 Configuration Tool manual 1 Select default unit by type 2 Then click on Read from the device 1 ...

Страница 16: ...tab displays product information totalizer information and Bluetooth data 5 The Loop tab select or confirm your 4 to 20 mA set point set your current alarm condition and type of output mode Note The 0252 Configuration Tool will be unable to connect to sensor when set to frequency ...

Страница 17: ...onfiguration 0252 Tool 7 8 8 The Calibration tab allows custom calibration via method of rate volumetric and zero flow calibration 7 The Monitor tab can graph or log the information to your local drive via file type with the extension CSV ...

Страница 18: ...GF Config Tool or 0252 Configuration Tool 2 Check wiring Use the wiring diagram picture in the manual Frequency Digital or Current Output is Erratic 1 Electrical Noise interference with the flow measurement 2 Possible air pockets traveling through the piping system 3 Pipe is not full and water flow creates splashing of the electrodes 4 Excess turbulence in fluid flow profile 1 Verify the grounding of the...

Страница 19: ...e elimination method if more Bluetooth units present Record the FlowtraMag s IDs or delete unused sensors 3 Restart GF Config Tool App 4 to 20 mA output is incorrect 1 The 4 to 20 mA in the FlowtraMag is not scaled properly 2 The Instrument used with the FlowtraMag has the 4 to 20 mA input not matching the FlowtraMag 3 The setting and or wiring for active passive 4 to 20 mA is done incorrectly 4 De...

Страница 20: ... Ordering Information Mfr Part No Code Description 3 2580 P T 010 159 001 874 FlowtraMag Blind PVC Union FKM 1 in 3 2580 P T 020 159 001 875 FlowtraMag Blind PVC Union FKM 2 in 3 2580 P T 040 159 001 876 FlowtraMag Blind PVC Flange FKM 4 in Accessories 3 0252 159 001 808 0252 Configuration Tool 854 040 4 inch SCH 80 Van Stone Flange 37X002009 4 inch EPDM Full Face Flange Gasket 150 ANSI bolt patter...