background image

2

Signet 2350 Temperature Sensor

1.  Prior to installation or removal:

• Depressurize and vent system
• Drain below sensor level 

2. Con

fi

 rm chemical compatibility before use.

3.  Do not exceed maximum temperature/pressure speci

fi

 cations.

4.  Wear safety goggles or faceshield during installation/service.
5.  Do not alter product construction.
6.  Dispose of properly; DO NOT INCINERATE!

General

Compatibility ......................... Signet 8900 Multi-Parameter Controller

Signet 9900 Transmitter
9950 Dual Channel Transmitter

Wetted material .................... PVDF

Measurement range

In-line installation .................. -10 °C to 100 °C (14 °F to 212 °F)
Submersible installation........ -10 °C to 85 °C (14 °F to 185 °F)
Response time, 

τ

 .................. 10  seconds

Process connection .............. 3/4 in. NPT male thread
Rear connection ................... 3/4 in. NPT male thread
Cable type ............................ 3 cond + shield, 22 AWG
Standard cable length........... 4.6 m (15 ft)
Shipping weight .................... 0.22 kg (0.5 lb)

Electrical

Power requirements
 (S

3

L) models ..................... 5 VDC ±10%, <1.5 mA

  4 to 20 mA models ............ 12-24 VDC ±10%
Short circuit & reverse polarity protected
Digital (S

3

L) output ................ Serial ASCII, TTL level 9600 bps

  Accuracy ........................... ±0.5 °C (±0.9 °F)
  Repeatability ..................... ±0.1 °C (±0.2 °F)
  Resolution ......................... 0.01 °C (0.02 °F)
  Update rate ....................... <100 ms
4 to 20 mA output
 Accuracy 

........................... ±32 

μ

A

  Repeatability ..................... ±0.1 °C (±0.2 °F)
 Resolution 

......................... <5 

μ

A

  Factory Span..................... 0 °C to 100 °C, 

fi

 eld-scaleable.

  Max loop impedance ......... 50 

Ω

 @ 12 V

325 

Ω

 @ 18 V

600 

Ω

 @ 24 V

  Update rate ....................... <100 ms

Environmental

  Relative humidity............... 0  to 95% (Non-condensing)
  Storage temperature ......... -55 °C to 100 °C (-67 °F to 212 °F)

Standards & Approvals

 CE
 RoHS 

Compliant

   Manufactured under ISO 9001 for Quality, ISO 14001 

for Environmental Management and OHSAS 18001 for 
Occupational Health and Safety.

  China RoHS (Go to www.gfsignet.com for details)
   Declaration of Conformity according to FCC Part 15. 

This device complies with Part 15 of the FCC rules. 
Operation is subject to the following two conditions: 
(1) This device may not cause harmful interference, and, 
(2)  This device must accept any interference received, 

including interference that may cause undesired operation.

Caution / Warning / Danger

Indicates a potential hazard. Failure to follow all warnings 
may lead to equipment damage, injury, or death

Electrostatic Discharge (ESD) / Electrocution Danger

Alerts user to risk of potential damage to product by ESD, 
and/or risk of potential of injury or death via electrocution.

Personal Protective Equipment (PPE)

Always utilize the most appropriate PPE during 
installation and service of Signet products.

Pressurized System Warning

Sensor may be under pressure, take caution to vent 
system prior to installation or removal. Failure to do so 
may result in equipment damage and/or serious injury.

Hand Tighten Only

Overtightening may permanently damage product threads 
and lead to failure of the retaining nut.

Do Not Use Tools

Use of tool(s) may damage product beyond repair and 
potentially void product warranty.

Refer to your local Georg Fischer Sales of

fi

 ce for the most 

current warranty statement.

All warranty and non-warranty repairs being returned must 
include a fully completed Service Form and goods must be 
returned to your local GF Sales of

fi

 ce or distributor. 

Product returned without a Service Form may not be 
warranty replaced or repaired.

Signet products with limited shelf-life (e.g. pH, ORP, chlorine 
electrodes, calibration solutions; e.g. pH buffers, turbidity 
standards or other solutions) are warranted out of box but not 
warranted against any damage, due to process or application 
failures (e.g. high temperature, chemical poisoning, dry-out) or 
mishandling (e.g. broken glass, damaged membrane, freezing 
and/or extreme temperatures).

Thank you for purchasing the Signet line of Georg Fischer 
measurement products.
If you would like to register your product(s), you can now 
register online in one of the following ways:
 •  

Visit our website 

www.gfsignet.com. 

Under 

Service and Support

 click on 

Product Registration Form

  •  If this is a pdf manual (digital copy), 

click here

Warranty Information

Product Registration

Safety Information

Specifi cations

Содержание Signet 2350

Страница 1: ... NPT allow submersion in process vessels or in line installation with conduit connection Integral adapters sold separately may be used to create a compact assembly with field mount style of the 9900 Transmitter 3 2350 090 1 Rev 11 10 17 3 2350 090 1 Signet 2350 Temperature Sensor English Description Dimensions English Deutsch Français Español Product Manual Pressure Temperature Graph 109 mm 4 27 i...

Страница 2: ...ccept any interference received including interference that may cause undesired operation Caution Warning Danger Indicates a potential hazard Failure to follow all warnings may lead to equipment damage injury or death Electrostatic Discharge ESD Electrocution Danger Alerts user to risk of potential damage to product by ESD and or risk of potential of injury or death via electrocution Personal Prot...

Страница 3: ...minals 3 9000 392 1 liquid tight connector in NPT To extend the wires longer than 4 6 m 15 ft Modify sensor 3 2350 1 or 3 2350 3 as described in figure 1 Terminate the three wires to the terminal board located in the cap assembly Add customer supplied wire to extend the cable Terminate to the transmitter or the 4 to 20 mA input device Apply sealant or PTFE tape to the process connection threads pe...

Страница 4: ...with 12 to 24 VDC Connecting the SHIELD to a direct Earth ground may reduce electrical noise interference Red wire is not used do not remove the heat shrink See 4 to 20 mA Span Adjustment page 5 Current Loop With No Junction Box Current Loop With Junction Box 3 8052 1 4 20 mA FREQ DATA WHT WHT BLU BLK N C RED SHIELD SENSOR INPUT SHIELD FREQ DATA White Black Power Supply DC 12 24 V Loop Input Loop ...

Страница 5: ...le 10 C to 85 C Be sure to allow sufficient time for the sensor to reach equilibrium to this temperature With power applied as described in on page 4 connect the RED wire to the BLACK wire for 15 seconds After about 10 seconds the output will rise to 22 mA and remain there until the RED wire is disconnected Disconnect the RED wire from the BLACK wire the 20 mA endpoint has been adjusted NOTE The o...

Страница 6: ...6 Signet 2350 Temperature Sensor Notes ...

Страница 7: ...7 Signet 2350 Temperature Sensor Notes ...

Страница 8: ...9000 392 2 159 000 841 Liquid tight connector kit PG13 5 1 piece 3 9900 396 159 001 701 Angle Adjustment Adapter Kit 5523 0322 159 000 761 Cable 3 conductor shield 22 AWG black red white shield 3 0252 159 001 808 Configuration Tool contact factory Custom cable length available Georg Fischer Signet LLC 3401 Aero Jet Avenue El Monte CA 91731 2882 U S A Tel 626 571 2770 Fax 626 573 2057 For Worldwide...

Отзывы: