16
John Bunn O Premier Oxygen Concentrator
6 tROuBlesHOOtIng guIde
syMPtOM
CHeCK / aCtIOn
Concentrator power is Check if concentrator is unplugged; plug in.
on and alarm is on, but
concentrator doesn’t
Check if circuit breaker is tripped; fix circuit
work
breaker.
If concentrator still doesn’t work, contact
graham-field equipment provider.
syMPtOM
CHeCK / aCtIOn
Concentrator works,
Inspect tubes and cannula to ensure that they are
but little or no oxygen
not kinked or blocked. smooth tubes to unblock.
from outlet
Check humidifier bottle, if so equipped, for leaks
or jamming. Give humidifier bottle a slight shake
to open safety valve. Clean humidifier bottle,
reinstall and tighten.
If condition persists, contact graham-field
equipment provider.
syMPtOM
CHeCK / aCtIOn
Flowmeter ball located Adjust flowmeter knob. If condition persists,
on “0” position or float contact Graham-Field equipment provider.
is unstable
syMPtOM
CHeCK / aCtIOn
Concentrator works,
If yellow light blinks, and alarm sounds at low
but makes abnormal
intervals, or yellow light illuminates continuously
sound
and alarm sounds, contact graham-field
equipment provider.
syMPtOM
CHeCK / aCtIOn
Concentrator works,
OK to continue use, but contact graham-field
but yellow fault light
equipment provider immediately. Concentrator’s
lluminates (JB0160-015/ oxygen concentration is safe, but decreasing.
JB0160-015-0)
When oxygen concentration is too low for safe
use, concentrator will shut off and optional O
sensor red warning light will illuminate.