2

GF 2751-7 pH Electronics

Caution / Warning / Danger

Indicates a potential hazard. Failure to follow all warnings 

may lead to equipment damage, injury, or death

Electrocution Danger

Alerts user to risk of 

potential of injury or death via 

electrocution.

Electrostatic Discharge (ESD)

Alerts user to risk of potential damage to product by ESD

Personal Protective Equipment (PPE)

Always utilize the most appropriate PPE during 

installation and service of GF products.

Note / Technical Notes

Highlights additional information or detailed procedure.

Refer to your local Georg Fischer Sales office for the most 

current warranty statement.

All warranty and non-warranty repairs being returned must 

include a fully completed Service Form and goods must be 

returned to your local GF Sales office or distributor. 

 

Product returned without a Service Form may not be  

warranty replaced or repaired.

GF products with limited shelf-life (e.g. pH, ORP, chlorine 

electrodes, calibration solutions; e.g. pH buffers, or other 

solutions) are warranted out of box but not warranted against 

any damage, due to process or application failures (e.g. high 

temperature, chemical poisoning, dry-out) or mishandling (e.g. 

broken glass, damaged membrane, freezing and/or extreme 

temperatures).

•  Remove electrical power from the transmitter before wiring 

input connections.

• 

Follow instructions carefully to avoid personal injury or 

damage to the electronics.

Warranty Information

Safety Information

Dimensions

Specifications

  61.0 mm

(2.40 in.)

  94 mm

(3.69 in.)

General

Compatible Electrode ..........2724-00 Flat pH Electrode

Compatible Instrument ........9950-X Chlorine Controller

Mounting ..............................DryLoc

®

 connection

Materials ..............................PC+PBT

Cable ...................................4.6 m (15 ft) 3 conductor shielded,

                                             22 AWG

Shipping Weight ..................0.64 kg (1.41 lb)

Performance

Electronics Accuracy ...........± 0.02 pH @ 25 °C

pH........................................ ≤ 0.01 pH

Response Time....................< 6 s for 95% of change

                                             (includes electrode response)

Operational Range ..............-1.0 to 15.00 pH

Electrical

Input Specifications

Input Impedance ..................> 10

11

 Ω

Temperature Drift .................± 0.002 pH per ºC

Input Resolution...................0.02 pH, 0.3 ºC

Output Specifications

Digital (S³L)..........................Serial ASCII, TTL level 9600 bps

Max. Cable Length ..............30 m (100 ft)

Environmental

Storage Temperature ...........–20 °C to 85 °C (–4 °F to 185 °F)

Operating Temperature ........0 °C to 85 °C (32 °F to 185 °F) 

                                              (electronics only)

Relative Humidity................. 0 to 95%, non-condensing

Enclosure Requirements .......NEMA 4X/IP65 with electrode 

                                             connected

Standards and Approvals

• 

CE, WEEE

• 

RoHS Compliant

• 

Manufactured under ISO 9001, ISO 14001 and ISO 45001

 

   China RoHS (Go to gfsignet.com for details)

          Declaration of Conformity according to FCC Part 15 

 

         

This device complies with Part 15 of the FCC rules. 

         Operation is subject to the following two conditions: 

 

         (1) This device may not cause harmful interference, and 

         (2) This device must accept any interference received,

 

         including interference that may cause undesired operation.