G E T IN G E S AF E S T EP AT P M O NI TO R

G E T IN G E S AF E S T EP AT P M O NI TO R

25

26

Error Code

Possible Causes

E1

Not applicable on the Getinge SafeStep 

ATP Monitor unit

E2

Temperature out of 

range

e The unit is being used outside of the 

     specified operating temperature range 

     (see section 10)

e The unit has been stored in an 

     environment which is outside of its 

     specified operating temperature range  

     –allow unit to acclimatise before use

f Unit damaged or faulty

E3

Erratic 

measurement

e Unit environment unstable or used in 

     an area of high electromagnetic noise

e Unit tilted whilst measurement being 

     performed

e Unit lid damaged and allowing light in

e Protective pocket dirty or severely 

     scratched

f Unit damaged or faulty

E4

User settings 

undefined

e The user configurable settings are  

     undefined - check and re-set the 

     CLOCK format, USER#, PLAN# and 

     PROG# data

E5

Program or User ID 

undefined

e Program thresholds not defined, and 

     have been reset to the default values 

     (100A and 100B)

e User ID not defined

E6

Calibration self-

checks failed

e Unit operating environment unstable 

e Protective pocket dirty or severely 

     scratched

e Lid not fully closed

f Lid seal damaged

f Protective pocket damaged

f Unit damaged or fault

E7

Internal memory 

failure

e Batteries are flat or loose

f Unit’s memory damaged or faulty

E8

Internal reader fault

e Batteries are flat or loose

f Unit’s sample reader is damaged or 

    faulty

E9

Internal error

e Batteries are flat or loose

e Unit dropped or subjected to shock or 

     vibration

f Unit’s damaged or faulty

The Supplier warrants the Getinge SafeStep ATP Monitor 
unit, when purchased new, to be free from defects in 
materials and workmanship, and will repair or replace, at 
their discretion, any SafeStep ATP Monitor unit which, used 
under proper conditions, exhibits such defects.

Under the terms of this warranty, the product must 
be returned in the original packaging, transportation 
prepaid, with a copy of the Proof of Purchase, to your local 
distributor.

Contact Getinge USA, Inc to receive authorisation to return 
the instrument, and enclose a detailed description of the 
problem.

8.1 Warranty Duration

This warranty is provided to the original purchaser for one 
year from the date of purchase.

In no event will Getinge USA, Inc be liable for indirect, 
incidental or consequential damages; the original user’s 
remedies being limited to repair or replacement of the unit 
at the manufacturer’s option.

8.2 Particular Exclusion

Unauthorised modification of any part of the SafeStep ATP 
Monitor unit or the use or attachment of any peripheral 
not supplied or specified by Getinge USA, Inc will void this 
Warranty.

WARNING: Use only the accessories and consumables 
supplied by Getinge USA, Inc. The use of any non Getinge 
USA, Inc supplied accessories and consumables will 
invalidate the warranty.

8. Unit Warranty and Returns

Содержание SafeStep

Страница 1: ...Getinge SafeStep ATP Monitor Operator s Manual ...

Страница 2: ...ling both during normal use and during storage and transportation CAUTION To avoid the risk of injury or choking never use the unit neck strap in situations where it could get caught in machinery etc Operating Precautions and Limitations of Use Batteries WARNING Use only non rechargeable alkaline batteries or rechargeable NiMH or NiCD batteries of types specified in section 10 WARNING Do not use b...

Страница 3: ... EEC and 92 31 EEC as realigned by 93 69 EEC Regulatory Limitations of Use 1 INTRODUCTION 1 1 Key Unit Features 1 2 Accessories and Consumables 1 3 Principle of Operation 2 BASIC UNIT OPERATION 2 1 Unit Description 2 2 Keypad Symbols 2 3 Display Layout and Icons 2 4 Fitting the Batteries 2 5 Turning On the Unit 2 6 Low Battery Indicator 2 7 Internal Self Calibration 2 8 Turning the Unit Of 2 9 Pow...

Страница 4: ...grammable users RS232 interface for results uploading to a PC Uses two standard or rechargeable batteries Replaceable protective sample pocket 1 Introduction 1 2 Accessories and Consumables Please contact your Getinge representative for details of unit accessories and consumables 1 3 Principle of Operation The Test Swab device uses a bioluminescent chemistry technology to convert an invisible conc...

Страница 5: ...s and their definitions Low battery warning icon Lid icon flashes when the lid needs to be closed or the RS232 connector cover needs to be properly tucked under the lid Insert icon arrow flashes when the testing device should be inserted Remove icon arrow flashes when an testing device needs to be removed Pass result icon Caution result icon Fail result icon Sample measurement reading in Relative ...

Страница 6: ... check sequence there must not be a Test Swab device in the unit and the lid must remain closed If the icon is shown with the arrow flashing open the lid and remove the Test Swab device from the unit If the icon is flashing close the lid TIP The Menu Select button can be pressed at any time to access the set up menu options Refer to section 4 for further details When the calibration self checks ar...

Страница 7: ...on and Kit Inserts for full details of how to use the Test Swab and Test Sponge device NOTE Always allow the unit sufficient time to acclimatise to any change in environment temperature or humidity before taking sample measurements To perform a sample measurement follow the steps below For Flat Surfaces Using the Test Swab Swab the sample area and activate the Test Swab device see Kit Insert for c...

Страница 8: ... button except the power button is pressed the countdown will restart To change the Standby Mode setting select the STANDBY MODE menu option by pressing the button Scroll to STANDBY MODE and press OK button Then use the up or down button to increase or decrease the Standby Mode duration followed by the OK button to store the new setting Standby Mode duration can be from 5 to 60 minutes in incremen...

Страница 9: ...results a Program location can be selected for retesting by pressing the P button see section 4 2 With the unit turned on the Set up Menu can be accessed by pressing the button Use the up and down buttons to scroll through the menu list and then press the OK button to select the required menu option or the button the exit the menu The following 11 menu options are then available User ID LCD Contra...

Страница 10: ...gured for either European format select D M Y for day month year or American format select M D Y for month day year D M Y or M D Y date D M Y or month M D Y month D M Y or date M D Y year TIP Pressing the button at any point will exit the clock set up mode leaving the time and date unchanged NOTE The clock does not have automatic daylight saving adjustment If this is required the time must be manu...

Страница 11: ... unit 5 1 Conditions of Use As Getinge USA Inc has no control over the specification state or use of any computer equipment on which this software is installed or used the following Conditions of Use apply IMPORTANT The software is provided as is without warranties of any kind either expressed or implied including warranties of merchantability or fitness for a particular purpose Getinge USA Inc sh...

Страница 12: ...he cable to your PC s USB port Step 9 Double click this icon to start the software Step 10 The first time you run the software the default user name Admin and password 1234 will be needed and may be changed in the software users form This is the default Admin account so the password should be changed if you choose so Step 11 Hit the Enter Key after you enter your login credentials If your ATP devi...

Страница 13: ...reports and manage users and hand held devices You must be an admin to manage devices and users All users can perform upload and download functions You can Upload and Download data to and from the handheld to the database for reporting needs Click the upload button in the menu bar with the ATP device connected a With the handheld connected choose the Upload To Display button to see which program p...

Страница 14: ...it using a wet cloth or by washing it under running water CAUTION Do not use solvents or other strong cleaning solutions as these may attack and deform the unit s plastic components and seriously degrade its performance 6 2 Replacing the Batteries For best results the batteries should replaced when the low battery warning icon is flashing Refer to section 2 4 for how to fit new batteries taking ca...

Страница 15: ...technical assistance giving full details of the error code This section lists typical problems that might be encountered when using the unit and their possible causes Some problems and causes can be rectified by the operator while others may require technical assistant Severity Action Required e This indicates a cause which can be rectified by the operator f This indicates a cause which may requir...

Страница 16: ...eader fault e Batteries are flat or loose f Unit s sample reader is damaged or faulty E9 Internal error e Batteries are flat or loose e Unit dropped or subjected to shock or vibration f Unit s damaged or faulty The Supplier warrants the Getinge SafeStep ATP Monitor unit when purchased new to be free from defects in materials and workmanship and will repair or replace at their discretion any SafeSt...

Страница 17: ... Bus used to connect computer peripherals to a PC Unit The Getinge SafeStep ATP Monitor unit General Unit dimensions W x H x D 72mm x 191mm x 32mm Unit weight including batteries approx 260g Operating temperature range Relative Humidity range 5 C to 40 C 20 85 non condensing Storage temperature range Relative Humidity Range 10 C to 40 C 20 95 non condensing Unit Details Measurement range 0 to 9999...

Страница 18: ...es in the United States or other countries Copyright 2017 Getinge AB or its subsidiaries or affiliates All rights reserved REVA Manufacturer Ruhof Corporation 393 Sagamore Ave Mineola NY 11501 1919 Sales Office Getinge 1777 East Henrietta Road Rochester NY 14623 www getinge com ...

Отзывы: