3 TECHNICAL DATA
Humidity
Max 80 % at 31 °C
Room temperature
5–40 °C
Weight
Approximately 100 kg
Internal dimensions
1600 x 495 x 375 (H x W x D)
Fan capacity
200 m³/h
Exhaust air
Fully open exhaust valve:
approximately 75 m³/h, max 90 °C.
Fully closed exhaust valve:
approximately 40 m³/h, max 90 °C.
Exhaust stub pipe Ø 100 mm valve
regulated.
Heat dissipation into the
room
Approximately 500 W if the exhaust
stub pipe is connected to the
exhaust ventilation.
Electrical
Max current
(A)
Power
(kW)
Max Pre
Fuse
200 V, 3+PE, 50/60 Hz
11.6 A
2.6 kW
C16 A
230 V, 3+PE, 50/60 Hz
13.3 A
3.4 kW
C16 A
240 V, 3+PE, 50/60 Hz
13.8 A
3.7 kW
C16 A
400 V, 3N+PE, 50/60 Hz
7.3 A
3.4 kW
C10 A
415 V, 3N+PE, 50/60 Hz
7.8 A
3.7 kW
C10 A
Pollution degree
2
Max altitude
2000 m
Indoor use only
Overvoltage category
2
Main supply voltage
fluctuations
+- 10% at nominal voltage short
term.
3 TECHNICAL DATA
8 (24)
USER MANUAL GETINGE DRYING CABINET S-363
<Doc_6001300802><Rev.C><Lang_en>