Contamination of the sterilised goods
“Danger of patient infection”
See Section 2.4, page 4, item 2.
)
Important
►
Only seal the packaging material when the red light
is flashing, ie when the sealing temperature has
been reached.
►
If you reduce the temperature, the red light goes out
(temperature too high). Wait until the temperature
adjusts to the new setting and the red light starts to
flash again.
►
If you raise the temperature, the red light stays on
continuously (without flashing) (temperature too low).
Wait until the temperature adjusts to the new setting
and the red light starts to flash again.
5.3 Setting
temperature
level 10 internally
The adjustment for temperature level 10 is on the underside of the case (1).
This is how to set the temperature:
−
Switch off the sealer and unplug it from the power supply.
GS-43-GB-002.doc