Troubleshooting
77811011
PhysioMem
®
PM 100 2G
39
Symptom
Cause
Recommendations
No display /
irregular
Display.
Malfunction possibly
caused by strong electro-
magnetic interference.
Switch off the device and
switch it on again, check use
environment.
8.2
Error Codes Displayed
E01-E05
If the display shows the GETEMED icon and one of the error codes
E01 to E05, a system error has occurred in the recorder. In this case,
contact the manufacturer's service or send the device to the manu-
facturer.
Fig. 20.
Example of error code display
E06
If the E06 error code is displayed, contact the technical service
E07
If the E07 error code is displayed, you have to restart the device.
Press the button (1) for 10 seconds until the display goes blank. Then
briefly press the button to restart the device.
If a restart does not solve the problem, contact the manufacturer's
service or send the device to the manufacturer.
E10
If the E10 error code is displayed, the operating temperature of the
device is too high or too low (see section 10.5). In this case, the de-
vice switches off automatically after 30 seconds. Only switch the de-
vice on again in an ambient where the temperature lays within the
temperature limits given in section 10.5.