Fig. 22
– ECG report, header area
The following information is displayed in the header of the ECG
report (Fig. 22):
(1)
– Device ID / serial number, device type, recording and trans-
mission time, event code (see table below).
The event code is displayed as a four-digit number:
Event
Value
Bradycardia, recorded automatically
0010
Tachycardia, recorded automatically
0020
Pause / asystole, recorded automatically
0040
Atrial fibrillation (AFib)
– start, recorded automatically
0080
Atrial fibrillation
– end, recorded automatically
0081
Bradycardia, recorded automatically during ongoing AFib
0090
Tachycardia, recorded automatically during ongoing AFib
00A0
Pause / asystole, recorded automatically during ongoing AFib 00C0
Time-triggered recording
1000
Time-triggered recording during ongoing AFib
1080
Manually recorded event
2000
Manually recorded event during ongoing AFib
2080
(2)
– Recorder setup, pre- and post-event times, event detection
limit values, detection of AFib and TIR ON/OFF