9
Fig.3-2 SD Card Calibration
3.3 Quality Control Calibration
To guarantee the accuracy of measurement and the comparability of data, calibration is
required before running patient samples. Please use corresponding SD card to calibrate
the analyzer before testing different batches of kits (Note: SD card for different batches
cannot be exchanged).
Steps:
Just attach the SD card to the SD card recognition zone, and the analyzer would
show the remind, click “OK” to import the reagent parameters (Fig.3-2).
Note:
User can also use insert the SD card into the SD card slot to do the calibration.
3.4 Sample Measurement
User can do sample detection as the following steps (Fig.3-3):
1) Edit patient’s information if needed.
2) Click “Start” after inserting the card, test item will be auto-recognized and the result will
be shown on the screen. Users can also see the voltage waveform by slide to the left
side.
3) Normally, the test card will auto-quit after testing, if not, please click on “Quit” icon to quit
manually.
Read parameter of hs-CRP+CRP, do you
want to use the data?
Remind
OK
Cancel
SD
card
or