12
Key
7
Conductivity electrode LRG 16-9
9
Measuring chamber, G ½, PN 40, 1.4301
b
Non-return valve
c
Ball valve G ½
d
Ball valve G
3
/
8
e
Measuring sensor ORG 12 / 22
f
Resistance thermometer Pt 100
g
Measuring electrode
Dimensions and examples of installation
- continued -
LRG 16-9
Important Notes
The conductivity electrode together with the measuring transducer ORGG 12 / 22 is
mounted in a bypass below the main line. This arrangement prevents gas bubbles and
flash steam from reaching the equipment. Please see also installation instructions for
OR 52-5 / OR 52-6.
Fig. 8
A throttling point, e. g. a non-return valve (GESTRA type RK 86) should be installed in the
main line.
To prevent air and dirt particles from getting into the bypass make sure that the bypass
is below the main line.
ALARM
1
2
ppm
sec
cal
0
Test
P
E
OIL
2bar
238°C
Tam
b
PN40
G1/2”
EST
RA
AG
G
·
TÜV.
WÜL.09-014/12-01
Münchener
LRG16-
9
GESTRA
G1/2 PN4 0
1.4301
Made in Germany
GESTRA
LRS 1-7
Conductivity electrode LRG 16-9 with measuring chamber for installation in a bypass together
with the measuring sensor ORG 12 / 22 of the oil & turbidity detector OR 52-5 / OR 52-6.
LRS 1-7
7
b
d
c
e
9
d
d
ORT 6