Add required amount of E2O

TM

 to water bottle (dose 

dependent on water bottle size). Proceed to fill water bottle 
with potable water (tap water or distilled water) to regular fill. 
Adding E2O

TM

 prior to filling bottle will reduce splatter and 

ensure product is properly mixed.

E2O

TM

 Shock Treatment

E2O

TM

 Shock Treatment | Directions

*

Please note the shock treatment is measured per chair,   

  regardless of the size of your water bottle. 
Shock treatment should be performed at the end of the day, 
after the last patient AND on a day when someone will be 
returning the following morning.

1. 

Empty any remaining solution from the water bottle.

2. 

Prepare shock treatment. Reference the appropriate 
number of chairs you need to shock from the 

Shock 

Treatment Dosage Chart 

and dilute only what you need, 

there is no wasted product. Add required amount of 
E2O

TM

 to empty water bottle, then dilute with potable 

water (tap water or distilled water) appropriately. 
Adding E2O

TM

 prior to filling bottle will reduce splatter 

and ensure product is properly mixed. 
As the shock treatment requires such a small amount of 
product per chair, we recommend one of the following: 

i. 

Mix enough for all chairs in one bottle, and then 
divide appropriately between the bottles in each 
operatory OR

ii. 

Mix enough for all chairs in one bottle, 
then travel to each operatory and purge the 
appropriate amount of shock treatment through 
the lines.

If the bottle is not completely full with shock treatment, 
shake well to ensure the solution comes in contact with 
all the interior surfaces of the bottle.  

3. 

Reattach bottle. 

4. 

Ensure no attachments are on the lines (handpieces, 
air/water syringes, ultrasonic and/or piezo scalers).

E2O

TM

 Shock Treatment | Dosage Chart

E2O

TM

 Daily Maintenance Treatment  

Directions

700mL water bottle: 1 pump E2O

TM

2L water bottle: 3 pumps E2O

TM

Number of Chairs

E2O

TM

Water (mL)

1

1 pump

100–150

2

2 pumps

200–250

3

3 pumps

300–350

4 pumps

400–450

5

5 pumps

500–550

6

6 pumps

600–650

7

7 pumps

700–750

8

8 pumps

800–850

9

9 pumps

900–950

10

10 pumps

1000–1050

11

11 pumps

1100–1150

12

12 pumps

1200–1250

13

13 pumps

1300–1350

14

14 pumps

1400–1450

15

15 pumps 

1500–1550

(100–150mL)  

water

Per Chair

1 pump E2O

TM

5. 

Place the lines into a sink and purge the lines for 2–3 
minutes. This ensures that the shock treatment has 
passed throughout the lines. Leave the shock treatment 
in the lines overnight.

6. 

In the morning remove and empty the bottle containing 
the shock treatment. 

7. 

Fill the bottle with E2O

TM

 

Daily Maintenance Treatment

Reattach bottle to unit. 

8. 

As per current standards/guidelines, purge the lines into 
a sink, with no attachments, for 2–3 minutes. Removing 
the attachments is an important step, as you do not 
want any debris from this treatment lodging in the 
attachments. 

*

Continue to use the E2O

TM

 daily maintenance treatment,  

refilling as needed. 

*

E2O

TM

 daily maintenance treatment/E2O

TM

 shock treatment 

maintains efficacy for up to 2 weeks when diluted.

 

A higher concentration of E2O

TM

 than the daily 

maintenance treatment.

 

When to perform a shock treatment:

1. 

Before initial use of E2O

TM

 to clean out your system.

2. 

Once a month; or as needed (if testing indicates 

bacteria above acceptable levels). 

3. 

When the unit has not been in use for 4 or more 

days AND the lines are left without water and 

waterline treatment.

Best Practices

 

All waterlines should be purged at the beginning of 
each workday by flushing them thoroughly with water 
for at least 2 to 3 minutes. Before purging is carried 
out, handpieces, air/water syringes and ultrasonic 
scalers should be removed from the waterlines.

 

Handpieces using water coolant should be run for 20 
to 30 seconds after patient care in order to purge all 
potentially contaminated air and water. The handpiece 
should then be removed and, following cleaning 
and disinfection of clinical contact surfaces, another 
sterilized handpiece is to be attached for use with the 
next patient. 

 

Always follow manufacturer’s instructions for use.

 

Complete dental unit waterline care requires not only 
continuous waterline maintenance, but also shock 
treatments and monitoring. 

 

When replacing the water bottle: Try to avoid touching 
the pick up tube. Always wipe the pickup tube with a 
clean cloth before placing it back into the bottle. 

 

Staff compliance of waterline protocol. 

TM

R1

M

O

N

IT

O

R

      SH

O

C

K

   MAINTAIN

Complete 

Waterline 

Care