Trouble
Possible Causes
Remedy
Low velocity air
Low voltage on house current
Measure voltage – contact
certification company
Filter loaded
Replace filters – contact
certification company
Improper adjustment of blower
motor
Adjust to proper reading on
pressure gauge – contact
certification company
Obstruction in plenum
Remove obstruction
Leak between blower and
plenum
Repair leak – contact certification
company
High velocity air
Filter rupture
Replace Filters – Contact
certification company
Supply filter gasket leak
Find and repair leak – contact
certification company
Exhaust air outlet obstruction
Remove obstruction
Improper adjustment of blower
motor speed control
Adjust to proper reading on
pressure gauge – contact
certification company
Non-laminar airflow
Large object obstructing airflow
Remove item
Large leak in supply filter
Replace Filters – Contact
certification company
Содержание BZ Series
Страница 7: ...Horizontal Laminar Flow Workstation User Manual 7 Airflow Diagram ...
Страница 18: ...Horizontal Laminar Flow Workstation User Manual 18 Service Log Date Service Provided ...
Страница 20: ...Horizontal Laminar Flow Workstation User Manual Appendix A Line Drawings ...
Страница 25: ...Horizontal Laminar Flow Workstation User Manual Appendix B Wiring Diagrams ...
Страница 26: ......
Страница 27: ......
Страница 28: ......
Страница 29: ......
Страница 30: ......
Страница 31: ......
Страница 32: ......
Страница 33: ......
Страница 34: ...Horizontal Laminar Flow Workstation User Manual Appendix C Sensocon User Manual ...