Using Commands
The following table shows the commands associated with each configuration that must be sent to
optical probe.
Command Type
Possible options
Command
Answer
Read Serial
10 bytes digits or letters
AT+AD Serial?
<CR><LF>
Serial Number=OP-
735
<CR><LF>
Read Bluetooth Name
10 bytes digits or letters
AT+AD Name?
<CR><LF>
Name=OP-735
<CR><LF>
Set Bluetooth Name
10 bytes digits or letters
AT+AD Name=OP-
735
<CR><LF>
Name=OP-735
<CR><LF>
Read Optical Probe
Model
OP-735, OP-745
AT+AD Model?
<CR><LF>
Model = OP-
735
<CR><LF>
Read Software Version
AT+AD Software
Version?
<CR><LF>
Version:01.01.01
<CR>
<LF>
Read Automatic Power
Off Time (Minute)
1 to 9
AT+AD
AutoPowerOff?
<CR><LF>
AutoPowerOff=5
<CR><L
F>
Set Automatic Power
Off Time (Minute)
1 to 9
AT+AD
AutoPowerOff=3
<CR><LF>
AutoPowerOff=3
<CR><L
F>
Read Working Mode
TRANSPARENT, IEC
AT+AD Mode?
<CR><LF>
CommunicationMode=IE
C
<CR><LF>
Set Working Mode
TRANSPARENT, IEC
AT+AD Mode=IEC
<CR><LF>
CommunicationMode=IE
C
<CR><LF>
Read Optical Port Frame Note*
AT+AD Frame?
<CR><LF>
Frame=300,7,E,1
<CR><
LF>
Set Optical Port Frame
Note*
AT+AD
Frame=300,7,E,1
<CR><LF>
Frame=300,7,E,1
<CR><
LF>
(Table no. 1)
Note*:
Set up and collection of the optical port communication frame is as follows:
•
Baud rate, Number of Bit, Parity, Stop bit
The values for each parameter are as follows:
•
Baud rate: 300, 1200, 2400, 4800, 9600, 19200
•
Number of Bit: 7, 8
•
Parity: N (none), E (even), O (odd)
According to the above description, if we want to connect with the meter with baud rate: 4800
and number of bits: eight and parity: E with the last 1 bit, the command format will be as
follows:
➢
AT+AD Frame =4800,8,N,1<CR><LF>
Note*:
Generally, for meters that comply with IEC 62056-21 standard, the command should be:
➢
AT+AD Frame =300,7,E,1<CR><LF>
Note*:
In case any command other than the mentioned commands are sent to the optical probe,
“Invalid Command!” will be returned and if parameters are not set correct, “Invalid Parameters!”
will be returned.
Upgrading Firmware
The firmware can be upgraded in two ways:
•
Send the file through the windows-based application
•
Send the file through any program via Bluetooth that can send commands