Gerhardt Vapodest 10s Скачать руководство пользователя страница 6

6

 

V

apodest

 10

s

3.3. Front view

3 Description of the system

Fig. 3.3: Front view Vapodest 10

Structural components and parts

  1  Quick clamping device with wedge

  2  Digestion tube, Order No.: 6100

  3  PTFE-Inlet tubing, steam, Art.-Nr.: 11815

  4  Holder for steam inlet tubing

  5  Viton-cone, Order No.: 6470

  6  Clamping for glassware

  7  Screw cap GL 18 with silicone seal, 

    Order No.: 16602

  8  PTFE-Inlet tubing, NaOH, Order No.: 11819

  9  PP-distributor with PP-threaded joint

 10  Distribution head, glass, Order No.: 7472

 11  Screw cap GL 32 with silicone seal, 

    Order No.: 16604

 

 12  Distillation condenser, Order No.: 7673

 13  Screw cap GL 14 with plastic screw connection, 

    Order No.: 16601

 14  Keyboard, 

    Description see chapter  5.1. "Keyboard"

 15  Mains switch, green, illuminated, Order No.: 11750

 16  Distillate outlet tubing, silicone 8/12 mm

    Order No.: 22704

17  Ventilation valve, Order No.: 6474

18  Platform

19  Drip tray, Order No.: 50015

 **  Plexiglas protection door (not illustrated)

    Order No.: 19017 

17

1

2

3

4

5

6

7

8

9

10

11
12
13

16

18

19

14

15

Содержание Vapodest 10s

Страница 1: ...10s VAP 10 05 Nov 2008_GB Please read this manual carefully before you start operating the system Please observe the safety instructions of this manual marked with in order to avoid any dangers resul...

Страница 2: ...neral 8 4 2 Tubing connections 8 4 3 Mains connection 9 4 4 Setting into operation 9 5 Operation 10 5 1 Keyboard 10 5 2 Setting of timer 10 5 3 Priming the tubes 11 5 4 Test run 11 5 5 Distilling a sa...

Страница 3: ...ulations in place It is understood that the user has to observe the accident prevention regulations in place the general accident prevention regulations EU regulations or other country specific instru...

Страница 4: ...out 3 litre per distillation minute Cooling water pressure 0 5 bar Pump capacity Diaphragm pump H2 O 11 ca 30 ml min with water Diaphragm pump Reagent 0 ca 170 ml min with NaOH Store tank any size rec...

Страница 5: ...the following list 3 2 Parts list 1 x Distillation system Vapodest10 complete 1 x Water inlet tubing 10 17mm with connections of 1 2 inch and 3 4 inch 2m Order No 22802 3 x PVC tubing 8 12 2 m order...

Страница 6: ...rder No 11819 9 PP distributor with PP threaded joint 10 Distribution head glass Order No 7472 11 Screw cap GL32 with silicone seal Order No 16604 12 Distillation condenser Order No 7673 13 Screw cap...

Страница 7: ...s off in case of excessive current consumption 5 Switch to rinse the steam generator Also see Chapter 7 2 2 Cleaning of steam gene rator Fig 3 4 Rear view Vapodest 10 6 Excess pressure steam outlet 7...

Страница 8: ...or distilled water and sodium hydroxide below the bench Make sure that the set of tanks are not placed any higher than the distillation unit 4 2 Tubing connections When connecting the tubings please o...

Страница 9: ...e more than 10 then you have to contact your service engineer Make sure that the mains switch at the front of the instrument is in position 0 Connect the mains cable to the rear appliance plug first C...

Страница 10: ...isplay of status steam power Green Instrument is ready for operation about 2 minutes after turning it on 5 Display of status operating mode Red distillation is running 6 Switch Addition reagent Manual...

Страница 11: ...the digestion tube The tube is now primed 5 4 Test run Each day before starting to distill samples you should do a distillation wit hout sample Don t be disturbed by noises deriving from the approx 1...

Страница 12: ...with about 50 ml boric acid and place into position 3 Manually dilute the sulphuric acid digestion with the 5 fold amount of distilled H2 O 4 Insert the steam inlet tube into the digestion tube 5 Low...

Страница 13: ...chemicals one after the other get a blank value a replacement of the chemicals might be re commended Increase the amount of water Exchange the glass condenser Clean the tubes without detergent or rin...

Страница 14: ...e amount delivered by the NaOH pump about 4 ml should be delivered per second Increase of the amount of boric acid Increase of the amount of acid This problem only occurs with catalysts containing mer...

Страница 15: ...iglass protection door 19017 Door handle black 18251 Door hinges black 1 pair 18250 Rubber foot GF21 self adhesive 17945 Drip tray PP 50015 Water inlet tubing 10 17 2 m 22802 PVC tubing 4 7 2 m 22601...

Страница 16: ...eaned Turn off the instrument The water tank must be filled with a minimum of 5 litres distilled water Set switch for rinsing on S see ill 3 4 pos 7 Turn on instrument to start rinsing process If tank...

Страница 17: ...id as of March 23rd 2006 As of this date C Gerhardt will take back all instruments delivered after August 13th 2005 without charges and will dispose of them according to the regulations or Gerhardt wi...

Страница 18: ...18 Vapodest 10s...

Страница 19: ...19 Vapodest 10s...

Страница 20: ...20 Vapodest 10s...

Отзывы: