Instructions for Use
Diffustik / Variant PFTstik
Serial no.: xx|8|401|yyy and 2401xxxxx
Page 80
Version: 8 | Release date: 16. August 2021
9.2.3
Ergoflow / Flow Sensor
Attention
Diffustik resp. the variant PFTstik could be damaged.
Reason: destruction of the orifice. Therefore:
•
Do not clean the interior of the sensor, mechanical or
with hard water jet!
•
Do not use any disinfectants which contain high
chloride concentrations!
1.
Remove the flow sensor
(in reverse order as in chap.
6.2.3. "Inserting the Flow Sensor"
described)
2.
Visible contamination wipe off
with a soft cloth
3.
Put the sensor in disinfectant
fluid
4.
Wash the sensor in distilled
water thoroughly
so that no residual dirt and
disinfectants remain
5.
Sensor air-dry until no moisture
is visible