User Manual - P series
22
www.georgin.com
VII.
Maintenance
1. Inspection frequency
Inspection frequency depends on several criteria:
»
The mechanical and/or electrical working life of the microswitch
»
The mechanical working life of the sensing element
»
Frequency and conditions of operation of the instrument
Georgin's standard* recommendation is to proceed as indicated:
1. Visual inspection of the instrument
(See Chapter II. Before installation)
2. Calibration
1. Visual inspection
2. Check cover seal
3. Checking / adjusting the calibration
0
1
6
12
Years
1. Visual inspection
2. Check cover seal
3. Checking / adjusting the calibration
1. Replacement to
be considered after
complete inspection
»
Year 0 :
1. Check that the paint has not been damaged, and check that there are no signs of deformation or impact
on the housing.
2. Calibration:
If the instrument is supplied adjusted: Check that the calibration is compliant, and readjust if necessary.
If the instrument is supplied unadjusted: Calibrate it according to the procedures given in this manual.
»
Year 1 :
After one year of standard operation, the mechanism can be considered to be run in. The following points
must therefore be checked:
1. Check that the paint has not been damaged, and check that there are no signs of deformation or impact
on the housing.
2. Open the cover and check that the cover seal is not damaged (cracks, etc.)
3. Check the calibration and adjust if necessary (applying the procedures given in this manual).
»
Year 6 :
After six years of standard operation, perform another inspection to ensure that the instrument is ageing well.
1. Check that the paint has not been damaged, and check that there are no signs of deformation or impact
on the housing.
2. Open the cover and check that the cover seal is not damaged (cracks, etc.)
3. Check the calibration and adjust if necessary (applying the procedures given in this manual).